2.0 Installation
2.3
Fitting the Heating Elements
Please refer to the Element Installation and Replacement in section 8.0.
Wear eye protection when handling the heating elements.
The molybdenum disilicide elements are EXCEPTIONALLY FRAGILE and
are packed separately, together with other items shown in the list.
items packed
quantity
Elements
6
Element Clamps
12
Element Clips
12
Braids
1 set
Separators/ Block
6
Chimney Unit
1
Remove the top access panel to allow element fitting.
Fit the heating elements vertically into the chamber. Handle the elements with great
care. The installation of these elements is described in section 0.1. Connect the
aluminium braids securely as indicated in the element installation instructions.
2.4
Electrical Connections
Connection by a qualified electrician is recommended.
This model is designed for connection to a three phase power supply. This may be with
neutral or delta, but a product must only be connected to the type and voltage of supply
for which it was made.
Check the product rating label before connection. The supply voltage should agree with
the voltage on the label and the supply capacity should be sufficient for the current on
the label.
The supply should be fused at the next size equal to, or higher than the current on the
label. A table of the most common fuse ratings is also given towards the back of this
manual. When the mains cable is factory fitted, internal fuses are also fitted. It is
essential that the operator ensures that the product is correctly fused.
Products with a factory fitted supply cable are designed to be wired directly to an
isolator or fitted with a line plug.
Products without a factory fitted supply cable require a permanent connection to a fused
and isolated supply. The product's electrical access panel should be temporarily
removed, and connections made to the internal terminals.
If the product is to be connected by line plug. The plug should be within reach of the
operator and should be easy to remove.
10
Содержание CARBOLITE GERO 3508 Series
Страница 7: ...Caution Double Pole Neutral Fusing 7 1 0 Symbols and Warnings ...
Страница 37: ...Power Limit View 37 4 0 3508 Dual Loop Cascade ...
Страница 57: ...Element Connections Example 3 phase 57 8 0 Repairs and Replacements ...
Страница 63: ...Notes Service Record Engineer Name Date Record of Work ...