background image

VentecLife.com

1

High Flow Nasal Cannula Kit Setup

The High Flow Nasal Cannula Kit is designed for use with all VOCSN configurations to deliver High Flow therapy 
to a patient. For more information about VOCSN, see the VOCSN instructions for use, available at 

VentecLife.com/

VOCSNmanual

. Read and follow the instructions for use provided with each kit component before use.

Note: Connecting incompatible parts to the Ventec One-Circuit may impair VOCSN performance. The organization 
responsible for setting up VOCSN, the Ventec One-Circuit, and any other components should ensure all connected 
parts are compatible with VOCSN before use. Use only Valveless, Heated Ventec One-Circuits to deliver High Flow 
therapy. 

To physically set up the items in the High Flow Nasal Cannula Kit:

Connect the Water Bag 

Pole to the Roll Stand by screwing it into place vertically. 

Attach the Humidifier Chamber to the Humidifier.

Slide the humidifier into place on the Roll Stand using the provided Humidifier bracket. Connect the Humidifier 
Temperature Probes and Heater Wire connections to the Humidifer as indicated in the Humidifier instructions. 

Connect the Humidifier Bypass to the VOCSN bacterial filter and the Humidifier Chamber.

Connect the Valveless, Heated Ventec One-Circuit to the Humidifier Bypass. Attach the Humidifier Temperature 
Probes and Heater Wire connections as indicated in the Humidifier and Ventec One-Circuit instructions for use.

Connect the Nasal Cannula to the Ventec One-Circuit. Read and follow the nasal cannula instructions for use.

Connect the O2 Hose to the high pressure O2 port in the back of VOCSN. Ensure it is tightly connected by 
twisting the connector clockwise until it stops.

6

1

4

3

Humidifier and Wiring

2

Humidifier

Chamber

7

O2

Hose

Water Bag 

Pole

Humidifier 

Bypass

Valveless,

Heated

Circuit

5

Nasal

Cannula