background image

INST

ALL

A

TION

   

Pre-Installation Instructions (Cont.)

VELOPEX INTRA-X

 

User Guide

•  Sitting of the VELOPEX

When using the machine in daylight or a darkroom, avoid sources of intense light. 
Do not mount the unit under a window, fluorescent light or flood lamp. 

IMPORTANT NOTE: A well ventilated position is mandatory.

The ambient temperature must be below 26ºC (78.8ºF).

NB: If the machine has been shipped or stored in excessive heat, allow time for it 
to cool to below 26ºC (78.8ºF) before installation. Temperatures above 57ºC will 
cause an E14 safety message.

Prevent sitting the machine above or near other electrical or mechanical equipment. 
Surfaces  susceptible  to  water  or  chemical  damage  should  be  avoided,  such  as 
carpeted areas.
     1. COUNTER

     2. ELECTRICAL SUPPLY

WARNING:  X-ray  radiation  can  be  harmful  to  patient,  technician  and  dentist. 
Inadequate lead shielding of the darkroom or film storage area will also cause 
fogging  from  exposure  of  films  to  stray  x-ray  radiation.  Consult  your  local 
codes, Health Department or Dental Equipment Dealer for proper construction 
of darkroom or placement of film processing equipment in the vicinity of x-ray 
radiation sources.

NOTE: Pictures are for instruction only and are not necessarily identical to product.
 

a. Use a stable and level counter that will support a weight of at least 50 Kg 
(100lbs).
b. When the machine is filled with chemicals, make sure the stand does not 
rock or shake. NEVER move the machine with chemicals in the tanks.

a. Refer to specification table, page v.
b. The power source must be within one metre (3 feet) of the machine. It 
should be easily accessible for operation and maintenance.

Содержание Intra-X

Страница 1: ...Do not plug into the electrical mains power supply before reading this manual or before filling the machine with liquids Register your machine by returning the Warranty Card to enable full customer su...

Страница 2: ......

Страница 3: ...could be dangerous if incorrectly installed or maintained outside the guidelines set out in this manual and the warranty will be voided There are no operator serviceable parts inside the machine This...

Страница 4: ......

Страница 5: ...ii Machine Accessories and Extras Supplied 12 12 Silicone Grease 13 Film Collecting Tray 14 Cleaning Brush 15 Turning Tool 1 Velopex X Ray Film Processor Left Side 2 Film Collector 3 Endo Slide 4 On...

Страница 6: ...NING 2EFER TO ANUAL TTENTION 7ARNING IGH 6OLTAGE 3YMBOLS USED WITHIN ANUAL 4IP TTENTION 7ARNING LEEP TTENTION 3IGNAL III Table of Symbols Symbol Description Alternating Current Off On Attention Warnin...

Страница 7: ...OADER NSTALLATION IF NOT INSTALLED ILLING WITH HEMICALS PERATION AINTENANCE ISPLAY RROR ODES 4ROUBLE 3HOOTING 3ERVICE OG AY EDIVANCE NSTRUMENTS IMITED LL RIGHTS RESERVED SSUE 4 34 4 0 2 4 4 42 5 3 54...

Страница 8: ...MM vPER MIN AX lLM WIDTH MM v 0ROCESSING TIME RY 7ET NDODONTIC MIN APPROX MIN APPROX NPUT 0OWER 7 NVIRONMENTAL ONDITIONS NDOOR USE ONLY NVIRONMENT 4EMPERATURE AXIMUM 2ELATIVE UMIDITY 5P TO ECREASING L...

Страница 9: ...rtant to have assistance Check contents of the box see page ii 1 Familiarise yourself with the layout of the machine by referring to the images as you progress through the manual 2 Lift the machine fr...

Страница 10: ...1 COUNTER 2 ELECTRICAL SUPPLY WARNING X ray radiation can be harmful to patient technician and dentist Inadequate lead shielding of the darkroom or film storage area will also cause fogging from expo...

Страница 11: ...rial and Modules TIP Start with the Dryer Module then the Water Module followed by the Fixer and Developer Modules Those should be removed by sliding the tanks out 2 3a 6 0 8 42 8 5SER UIDE 2 4 PERATI...

Страница 12: ...OVER 25 3ENSOR 4 0 4HE TIME TAKEN TO ACHIEVE THE CORRECT TEMPERATURE DEPENDS ON THE ROOM TEMPERATURE USUALLY MIN 7 2 EFORE PLACING IN THE MACHINE INTRA ORAL lLM PACKETS SHOULD BE WIPED CLEAN OF ALL M...

Страница 13: ...TH RETAINING SCREW AYLIGHT OADER NSTALLATION CONT 3LIDE OADER OVER ILM NTRY UIDE AND 25 3ENSOR 0RESS THE OADER lRMLY AGAINST ACHINE MAGES FOR INSTRUCTION 9 Two Slide Loader over Film Entry Guide Press...

Страница 14: ...IDE 0 2 4 PERATING THE 0ROCESSOR 0LUG IN ELECTRIC CORD AND SWITCH ON 4HE MACHINE WILL THEN RUN THROUGH A HIGH SPEED START UP CYCLE OF MINUTES 4 0 FTER APPROX TEN MINUTES THE PROCESSOR WILL GO INTO STA...

Страница 15: ...ES AND COOLING DOWN IS INDICATED IN THE LINES DESCENDING FROM LEFT TO RIGHT MAGES FOR INSTRUCTION 9 Fill the tanks up to the lowest edge of the Sight Glass LowerTransportModulescarefully into their re...

Страница 16: ...he lid open 6 0 8 42 8 5SER UIDE 0 2 4 PERATING THE 0ROCESSOR 4HE 6ELOPEX IS EQUIPPED WITH AUTOMATIC STAND BY MODE 4O INITIATE PROCESSING FEED lLM THROUGH NTRY 3LOT F THAT FAILS TO WORK SLIDE lNGER OV...

Страница 17: ...OR INSTRUCTION 9 6IEWING 7INDOW IS CLOSED WHEN OPERATING MACHINE MAGES FOR INSTRUCTION 9 Plug in electric cord and switch on The machine will then run through a high speed start up cycle WARNING Alway...

Страница 18: ...T NIGHT NCORRECT TEMPERATURE IS INDICATED WHEN THE MESSAGE ON THE DISPLAY READShT Ov 2EADY TO USE IS INDICATED BY THREE HORIZONTAL CENTRAL LINES AND COOLING DOWN IS INDICATED IN THE LINES DESCENDING F...

Страница 19: ...EADShT Ov 2EADY TO USE IS INDICATED BY THREE HORIZONTAL CENTRAL LINES AND COOLING DOWN IS INDICATED IN THE LINES DESCENDING FROM LEFT TO RIGHT 0 2 4 PERATING THE 0ROCESSOR 4HE 6ELOPEX IS EQUIPPED WITH...

Страница 20: ...re the film on the guide provided to post the film through the processor WARNING Make sure the film is fed in squarely and in the centre of the entry slot short edge leading 6 0 8 42 8 5SER UIDE 0 2 4...

Страница 21: ...IS INDICATED IN THE LINES DESCENDING FROM LEFT TO RIGHT MAGE FOR INSTRUCTION 9 6IEWING 7INDOW IS CLOSED WHEN OPERATING MACHINE MAGES FOR INSTRUCTION 9 It is suggested that the VISCHECK Quality Manage...

Страница 22: ...results use VELOPEX chemicals TIP Use the chart provided and keep a note of the date when chemicals were changed 5 6 6 0 8 42 8 5SER UIDE PERATING THE 0ROCESSOR 0LUG IN ELECTRIC CORD AND SWITCH ON 4H...

Страница 23: ...or sink filled with fresh cold water Add one Velopex Cleaning Tablet and soak for at least 10 minutes Rinse thoroughly with water to remove all cleaning solution TIP Add a little domestic fabric soft...

Страница 24: ...RRECT TEMPERATURE DEPENDS ON THE ROOM TEMPERATURE USUALLY MIN 0LUG IN ELECTRIC CORD AND SWITCH ON 4HE MACHINE WILL THEN RUN THROUGH A HIGH SPEED START UP CYCLE OF MINUTES 4 0 FTER APPROX TEN MINUTES T...

Страница 25: ...Drive motor overloaded 1 Switch off the Machine 2 Clean the Gears on the Modules 3 Switch on the Machine 4 If still E12 Call for Service E13 Transport motor Sensor Inactive or Modules seized 1 Switch...

Страница 26: ...HROUGH ILM NTRY UIDE ACHINE IS IN STANDBY MODE 3WIPE lNGER ACROSS THE 2UN SWITCH MACHINE MAY BE IN STANDBY MODE 4RANSPORT ODULES NOT IN 0LACE PEN MACHINE LID AND CHECK CORRECT ENGAGEMENT OF TRANSPORT...

Страница 27: ...NDER SPOTLIGHTS OR STRIP LIGHTS O NOT REMOVE YOUR HANDS OUT OF THE HAND ENTRY PORTS BEFORE THE lLM HAS FED COMPLETELY INTO THE MACHINE FOG AT ONE END ONLY OF THE lLM INDICATES PREMATURE REMOVAL OF HAN...

Страница 28: ...ES CALL FOR SERVICE ILM XPOSURE HECK lLMS HAVE BEEN EXPOSED CORRECTLY ILMS IRTY OR ARKED 7ATER 4ANK IRTY LEAN MODULE THOROUGHLY AND CHANGE WATER DAILY INSUFlCIENT CLEANING CAN LEAD TO A BUILD UP OF AL...

Страница 29: ...LM 7HITE ARKS IXER HECK FOR CONTAMINATED OR WRONGLY MIXED IXER IXER EPOSITS HECK FOR RYER ODULE CONTAMINATED BY lXER DEPOSITS LACK 0ARALLEL INES IRT LEAN ENTRY SLOT OR FEED GUIDES LEAN BRIDGE OVER ROL...

Страница 30: ...OFF AT THE AINS 0OWER 3WITCH FOR TWO MINUTES AND SWITCH IT ON AGAIN F THAT DOES NOT CORRECT THE FAULT CALL FOR SERVICE ONTAMINATION HEMICALS 7HEN CHANGING HEMICALS MAKE SURE THE TANKS ARE COMPLETELY E...

Страница 31: ...Service Log VELOPEX INTRA X 23 User Guide ATE 3ERVICE ESCRIPTION 3ERVICED Y Machine Installation 3ERVICE OG...

Страница 32: ...STED ENREGISTR ETL CONFORMS TO CONFORME UL 61010 1 THIS PRODUCT HAS BEEN TESTED TO THE REQUIREMENTS OF CAN CSA C22 2 No 61010 1 SECOND EDITION INCLUDING AMENDMENT 1 OR A LATER VERSION OF THE SAME STAN...

Отзывы: