14

Remove the AquaCare Twin from its box and retain the packaging, use to return unit 
for service as required.

If you are using the AquaCare Twin with the Velopex Patient Treatment Centre, place 
the machine on the top shelf of the trolley, following the instructions received with the 
trolley. See RECOMMENDED AND COMPLIMENTARY ACCESSORIES, Pg 39.

If you do not have a specialised trolley, place the machine securely on a 

rigid, sturdy 

desk

 or table top.

Do not place the machine close to, or allow it to hang over, the edge of the table top.
The machine is designed to have all the controls within easy reach of the dentist when 
situated in this way.

The unit should be kept within a temperature range of +15ºC and +40ºC, and should 
also be kept below 80% humidity.

Siting The AquaCare Twin

AquaCare Twin: Installation

Содержание AquaCare TWIN AT/100000

Страница 1: ...10 40 C 80 IIa Medical Device CLASS 3 066KG ATTENTION DO NOT CONNECT TO THE AIR SUPPLY BEFORE READING THIS GUIDE AquaCare Twin User Manual Iss 19 LIT0171P Issued 06 22 Installation Operation Maintenan...

Страница 2: ...L AQUACARE TWIN FEATURES 10 REAR VIEW OF AQUACARE TWIN 11 FOOT PEDAL 11 HANDPIECE 12 POWDER CARTRIDGES 13 INSTALLATION SITING THE AQUACARE TWIN 14 ATTACHING THE FOOT PEDAL 15 INSERTING THE FLUID CONTA...

Страница 3: ...ity Preparation Removal of Composite Restorations Bonding Orthodontic Brackets MAINTENANCE STERILISING THE HANDPIECE 31 HANDPIECE MAINTENANCE 32 TWIN TUBE ASSEMBLY MAINTENANCE 32 PURGING THE AQUACARE...

Страница 4: ...hing of enamel metal and porcelain Contraindications For Use The AquaCare Twin is not intended for removal of calculus The AquaCare Twin is not intended for cutting or removal of amalgam Please refer...

Страница 5: ...perating if coming from a cold environment ATTENTION Please read all the following important warnings before using the AquaCare Twin Safety Warnings AquaCare Twin Introduction The disposable feedline...

Страница 6: ...he performance The fluid used with the machine must be a biostatic solution like the AquaSol from Velopex see page 16 Site the foot pedal carefully so that it cannot be operated by accident Always clo...

Страница 7: ...tings 3 Bar 3 Bar Midway Position for Cutting media 0 1 2 3 4 Media gauges NOTE 1 The Power and Media controls operate from zero to full in one rotation I E Between and 2 The Fluid control has approxi...

Страница 8: ...AquaCare Twin is operated by pressing the foot pedal in three modes Dry Wash and Cut Clean The air supply power dosing chamber media volume and fluid volume can all be adjusted using the seven contro...

Страница 9: ...aSol H 1x Cartridge Red Coarse cutting powder 53 micron aluminium oxide I 1x Cartridge Blue Fine cutting powder 29 micron aluminium oxide J 1x Cartridge White cleaning polishing sodium bicarbonate K 1...

Страница 10: ...Powder Out Connection 13 Fluid Out Connection 14 Power Control Prep 15 Powder Chamber Selector 16 Power Control Prophy 17 Media Control Prophy 18 Media Gauge Prophy 19 Air Pressure Gauge Prophy 20 Air...

Страница 11: ...n Using The Foot Pedal Air 1 2 3 Fluid Air Powder Fluid Air 24 Foot Pedal Connection Black 4 Way Tubing 25 Foot Pedal Connection White 4 Way Tubing 26 Air In Supply Blue 27 Foot Pedal Connection Green...

Страница 12: ...Features 30 Fluid Connection 31 Feedline And Tip 32 Nozzle Range Of Sizes 33 Handpiece Front 34 Hand Grip 35 Powder Connection 36 Handle 37 Twin Tube 38 Handpiece Holder 39 Check Valve AquaSol PROPHY...

Страница 13: ...Chamber Lid Prep 44 Powder Cartridge Prep 45 Open Dosing Chamber Prep Aluminium Oxide 53 For cutting cavity preparation and composite preparation Aluminium Oxide 29 For cutting cavity preparation and...

Страница 14: ...NDED AND COMPLIMENTARY ACCESSORIES Pg 39 If you do not have a specialised trolley place the machine securely on a rigid sturdy desk or table top Do not place the machine close to or allow it to hang o...

Страница 15: ...dentist B Connect the foot pedal to the back of the machine by pushing each coloured tube into its colour corresponding connector WARNING Accidental operation of the foot pedal can be hazardous To avo...

Страница 16: ...uci n acuosa Contiene 17 5 w w de etanol Co ep e a e oe o ec o o ac x a ep a o Bo pac op co ep a 17 5 o o o c p a Information Holding the fluid container with the lid upward turn the container over an...

Страница 17: ...e Purging the AquaCare Pg 33 WARNING The handpiece should be sterilised before use and after each subsequent use See Maintenance Sterilising the Handpiece Pg 31 Feedline A See Parts and Features Handp...

Страница 18: ...machine by pushing it fully into the blue connector See page 40 regarding installing an inline filter Air Supply Requirements Clean dry air to ISO 8573 Class 1 4 1 Pressure between 4 bar 72psi and 7...

Страница 19: ...Small quantity of water Very fine mist SPRAY Larger quantity of water Heavier spray K Set the powder chamber selector to the prep position Press the foot pedal fully down to the CUT position 3 You sho...

Страница 20: ...ge down into the dosing chamber until it sits comfortably inside If the cartridge sits unevenly or above the top of the dosing chamber then lift it out and re insert Check that the rib of the powder c...

Страница 21: ...21 AquaCare Twin Installation Inserting The Powder Cartridges 1 2 1 2 I PUSH IN CLEAR TAG CLEAR TAG PULL BACK TO LOCK TAG PUSH IN WARNING DO NOT FULLY REMOVE CLEAR TAG PULL OUT 1 2 1 2 I...

Страница 22: ...ving the spray On a paper towel you should see a ring of white powder inside an outer ring of fluid The Media volume range as shown by the media gauge varies between Slightly under midway position Low...

Страница 23: ...y waste powder efficiently L Move the foot pedal between polish wash and dry positions to observe the results M Try using this technique to polish a drinks can The paint will be removed with no damage...

Страница 24: ...ip Its ability to cut wet or dry gives it a high flexibility of use Local anaesthetic can be offered to nervous patients however it is generally not required as fluid abrasion is painless in most case...

Страница 25: ...be used with the extraction cone sited below the patient s chin 7 IMPORTANT To avoid unnecessary damage when switching from Aluminium Oxide to Sodium Bicarbonate you must purge the powder See MAINTEN...

Страница 26: ...a wide particulate beam of high velocity and medium particulate density This is to prevent removal of the tooth enamel but provide sufficient power to remove staining The cutting nozzle should be hel...

Страница 27: ...tooth structure is revealed The beauty of the procedure is its speed and the absolute control possible ensuring that the minimum amount of tooth structure is removed The newly prepared surface being d...

Страница 28: ...se a slow circular motion over the affected area until sound tooth structure is revealed The tooth is now ready to be filled in the normal way For larger cavities a combination of the AquaCare Twin an...

Страница 29: ...the surface of the old composite restoration and about 2mm or 3mm from the surface The best approach is to use a slow circular motion over the old restoration beginning at the edge particularly at th...

Страница 30: ...e of the orthodontic bracket at the place of attachment This is achieved by holding the cutting nozzle 2mm above the surface of the tooth and gently moving it in a circular motion over the required ar...

Страница 31: ...all syringes with AquaSol solution push each on to the twin tube in turn and pump AquaSol solution through the tubes Should another solution be obtained by the user it must provide the required biolog...

Страница 32: ...autoclaved Wash the Twin Tube Assembly in mild soapy water and rinse thoroughly Allow to dry completely before use Always purge the tubes with Powder and Fluid thoroughly before use Inspection Ensure...

Страница 33: ...he media volume as desired We recommend that the fluid bottle is checked prior to surgery to ensure it will not run out whilst working Replace the container if the level is low prior to surgery Always...

Страница 34: ...e do not use the AquaCare unit until it has been inspected by a Velopex representative or Authorised Distributor These will also be checked during the annual service and safety check please see page 3...

Страница 35: ...he outside surfaces of the unit i e the Dosing Chambers Lids Control Knobs Foot Pedal with a lint free cloth lightly moistened with water or AquaSol Note See warning note regarding alcohol based wipes...

Страница 36: ...he case contact a Velopex Representative or Authorised Dealer who will be able to assess the damage to the AquaCare unit Symptom No air flow through the nozzle Action 1 Ensure that the power dial is t...

Страница 37: ...r Authorised Distributor for further help Symptom Irregular flow or no flow of cutting fluid to the tip Action 1 Check that the fluid volume control is not set to minimum Rotate clockwise to increase...

Страница 38: ...k from the machine Turn the unit off and carefully wipe the inner Lid seal and Dosing Chamber sealing faces Screw the lid on tightly until it has reached the end of its thread NOTE If the fault cannot...

Страница 39: ...chines in the range some of these tasks could have been performed by the end user however the latest MK5 design does not allow access to the machine and therefore must be performed via Velopex support...

Страница 40: ...Handpiece 0 6mm Silver Handpiece 0 8mm Gold 500 ml AquaCare Fluid Bottle AquaSol 500 ml Bottle Empty 500ml x 6 AquaCare Economy Pack Nozzle Cleaning Wire Nozzle Wash Syringe Pack of 2 Chalk Stick Part...

Страница 41: ...on of the machine In Order to give some indication of the incoming Air Quality we are now introducing a free of charge in Line filter that needs to be installed in the incoming air supply The filter s...

Страница 42: ...the end of the twin tube in to it Attach the twin tube to the cutting tip Make sure the narrow tube is attached to the narrow socket and the wider tube to the wider socket Push and connect the tubing...

Страница 43: ...43 Notes AquaCare Twin Notes...

Страница 44: ...lopex com 9051184 Dental Air Abrasion Polishing Unit Installation Operation Maintenance Manual AquaSol PROPHY PREP AquaCare Twin User Manual Iss 19 LIT0171P Issued 06 22 1639 TWIN AT 100000 ENGLISH TW...

Отзывы: