5.2. FUSE
REPLACEMENT
WARNING
To avoid electric shock, remove the test leads from the circuit under
test before replacing the fuse. For continued protection, only replace
fuses with specified ratings:
F1: F 200mA/250V F2: F 10A/250V.
5.2.1. Fuses rarely need replacing and almost always blow as a result of
operator error.
5.2.2. Loosen the screw fixing the case and remove it.
5.2.3. Replace the blown fuse with the specified ratings.
5.2.4. Put back the case.
5.3.
TEST LEAD REPLACEMENT
WARNING
Full compliance with safety standards can only be guaranteed if used
with the test leads supplied. If necessary, they must be replaced with
leads of the same type or ones with the same electrical ratings
Electrical ratings of the test leads: 1000V 10A
The test leads must be replaced if bare wire is exposed.
5.4. HOW TO USE THE BRACE
5.4.1. When the meter is not in use: attach the case to the panel face for
safekeeping.@@
5.4.2. When measuring: attach it either to the back of the case or use it as a
stand as illustrated in the cover. It has 3 different positions for easy
reading.
5.4.3. Attach the stand to the top part of the meter and use as a hook for
hanging.