![Velleman DVM601 Скачать руководство пользователя страница 8](http://html.mh-extra.com/html/velleman/dvm601/dvm601_user-manual_862154008.webp)
DVM601
22/08/2013
©2008 Velleman
®
Components nv
8
•
For continuity tests, connect the test leads to two points of the circuit to
be tested. If continuity exists (resistance < 50Ω), the built-in buzzer will
sound. If the circuit is open, “OL” will be indicated.
Non-contact AC voltage test
•
Set the rotary switch [3] to any position (except OFF).
•
The AC-sensor is located at the top of the meter, near the light sensor
[9]. Point the AC-sensor towards the suspected AC-source. When AC
voltage is present, the indicator [16] will light up. The stronger the AC-
source, the more the indicator [16] will be lit.
7.
Battery
•
When the low battery indication (
)
appears, replace the internal
batteries. Turn the rotary switch [3] to the OFF-position.
•
Always disconnect test leads when replacing the battery. Do not us the
device without batteries installed.
•
The battery cover is located on the back of the device and is closed with two
screws. Remove the screws and open the battery compartment.
•
Remove the battery and insert a new 9V (E-block) battery following the
polarity as indicated in the battery holder.
•
Reinstall the cover and secure it with the two screws.
•
Remove the battery when the device is not in use for a longer period of time to
avoid leakage.
•
Do not attempt to recharge non-rechargeable batteries, do not puncture and do
not throw in fire as they may explode.
WARNING: handle batteries with care, observe warnings on
battery casing. Dispose of batteries in accordance with local
regulations.
Keep batteries away from children.
8.
Fuses
•
To replace an internal fuse, turn the rotary switch [3] to the OFF-position.
•
Always disconnect test leads when replacing fuses. Do not use the device
without the proper fuses installed.
•
The access the fuses, the whole back cover must be removed. Loosen the 6
outer screws (2 are located under the foldable stand). Do not remove the
battery cover screws (see §7). Lift the whole back cover. Do not touch the
electronic circuitry inside.
•
Replace fuses only by new ones with following specifications:
•
Fuse 1 (large): F 10A/600V fast blow
•
Fuse 2 (small): F 500mA/660V fast blow
Using wrong fuses or short-circuiting fuse holders can lead to
potentially life-threatening situations.
•
Close the back cover and secure the 6 screws.