Velleman DVM345DI Скачать руководство пользователя страница 6

DVM345DI                                                                                                                                                                GB 

 
4.7. Diode Test 
 
1. Connect the black test lead to the "COM"-

MDFNDQGWKHUHGWHVWOHDGWRWKH9 

-jack. The 

polarity of the red test lead is positive (+). 

2. Place the function switch in the "

"-position. 

3. Connect the red test lead to the anode and the black test lead to the cathode of the diode to 

be tested. 

4. The forward voltage drop of the diode will be displayed on the LCD. 
 
4.8.  Transistor Test 
 
1. Place the function switch in the "hFE"-position. 
2. Determine whether the transistor is NPN or PNP and locate the emitter, base and collector 

leads. Insert the leads into the proper holes of the hFE-socket on the front panel. 

3. The display will show the approximate hFE-value at the moment of testing. Base current 

10µA, Vce 3.0V. 

 

 WARNING : 

 
Disconnect all test leads from the circuit to be tested before connecting the transistor with the 
hFE-socket. Make sure that no components are connected to the hFE-socket when performing 
voltage measurements with test leads. 
 
4.9.  Interfacing the Meter with a PC 
 
1. Run the RS232C-cable from the computer

's serial ports to the meter. 

2. Press "ON/OFF" to activate the meter and turn on the computer. 
 

 WARNING : 

 
1. Only use a RS232C serial interface cable. Do not attempt to modify the length of this cable. 
2. Please read the information on the supplied disk. 
 

5. SPECIFICATIONS 

 
Max. accuracy is achieved during a one-year period after calibration. Ideal circumstances 
require a temperature of 18 to 28°C (64 to 82°F) and a max. relative humidity of 75%. 
 
5.1. General Specifications 
 
Max. Voltage between 
Terminals and Earth Ground :  1000VDC or 750 Vrms AC (sine wave) 
Power Supply 

:  9V-battery (NEDA1604 or 6F22) 

Ranging Method 

:  Auto/Manual 

Display 

:  3 ¾ digit LCD, 38-segment bar graph 

Overrange Indication 

:  "OL" appears on display 

Polarity Indication 

:  "-" is displayed automatically 

Battery-Low Indication 

:  The "

"-symbol is displayed 

Operating Temperature 

:  5 to 35°C (41 to 95°F) 

Содержание DVM345DI

Страница 1: ...DVM345DI DIGITAL MULTIMETER DIGITALE MULTIMETER MULTIMETRE NUMERIQUE DIGITALES MULTIMETER User Manual Gebruikershandleiding Manuel d utilisation Gebrauchsanleitung ...

Страница 2: ...ed test leads If necessary they should be replaced with identical leads or leads with identical electric ratings All test leads should be in good working order 2 1 Use Never exceed the specified limit values for the various measurement ranges Do not touch unused terminals when the meter is connected to a circuit Place the range selector switch in the highest position if the value to be measured is...

Страница 3: ...audible continuity 3 1 ON OFF push button Used to turn the device ON or OFF The beeper sounds when this button is pushed 3 2 DC AC RU button 8VHG WR VHOHFW RU YROWDJH UDQJH RU FXUUHQW UDQJH RU range The beeper sounds when this button is pushed 3 3 R H button Press this button to switch to manual operation when the device is in the auto ranging mode R H will be displayed Hold the button for more th...

Страница 4: ...h R H for manual range selection 4 The measured value will be displayed on the LCD The polarity of the red test lead is indicated with DC measurements 4 2 Current Measurements 1 Connect the black test lead to the COM jack and the red test lead to the mA jack for measurements up to 400mA Switch the red test lead to the 10A jack for measurements up to 10A 2 Place the function switch in the 4mA 400mA...

Страница 5: ...ket WARNING Disconnect all test leads from the circuit to be tested before inserting the capacitor in the capacitance socket Make sure that no components are connected to this socket when performing voltage measurements with test leads 4 5 Temperature Measurements 1 Place the function switch in the TEMP position 2 Connect the K type thermocouple with the temperature socket on the front panel and c...

Страница 6: ... hFE socket Make sure that no components are connected to the hFE socket when performing voltage measurements with test leads 4 9 Interfacing the Meter with a PC 1 Run the RS232C cable from the computer s serial ports to the meter 2 Press ON OFF to activate the meter and turn on the computer WARNING 1 Only use a RS232C serial interface cable Do not attempt to modify the length of this cable 2 Plea...

Страница 7: ...n in rms of a sine wave Frequency range 40Hz 400Hz 5 4 Resistance Range Resolution Accuracy 400 0 1 4k 1 40k 10 400k 0 1k 4M 1k 1 2 of rdg 3 digits 40M 10k 3 0 of rdg 5 digits Max open circuit voltage 3 0V Overload protection 250VDC or 250Vrms AC for all ranges 5 5 DC Current Range Resolution Accuracy 4mA 1µA 1 2 of rdg 3 digits 400mA 0 1mA 1 2 of rdg 3 digits 10A 10mA 2 0 of rdg 8 digits Overload...

Страница 8: ...y 0 to 400 C 1 C 3 0 of rdg 3 digits 401 to 750 C 1 C 3 0 of rdg 5 digits 5 9 Audible Continuity Function Description Built in buzzer sounds if continuity exists 30Ω 5 10 Diode Function Resolution Test Current Open Circuit Voltage 1mV 25µA 3 0V 5 11 Transistor Function Range Base Current Vce hFE 1 to 1000 10µA 3 0V 6 ACCESSORIES User manual Set of test leads 9V battery NEDA1604 or 6F22 K type ther...

Отзывы: