DCM267                                                                                                                                                                    GB 

clothing and rubber-soled shoes. Put a rubber mat or approved insulating material under your 
feet. 

• 

Exercise extreme caution when working with voltages in excess of 60VDC or 30 Vrms AC. 
Keep your fingers behind the probe barriers while using the device. 

• 

Never use the device to measure voltages that (may) exceed the max. input value for a 
particular function. 

• 

Never touch wiring, connections or live circuits while the device is in use. 

 

3. Maintenance 

 

• 

Disconnect all test leads from energised circuits prior to opening the housing. 

• 

Do not use the device unless the cover is in place and securely fastened. 

• 

Do not apply abrasives or solvents to the meter. Use a damp cloth and a mild detergent for 
cleaning purposes. 

• 

Only qualified technicians should attempt to repair or calibrate this device. 

• 

If you insist on calibrating the instrument yourself, you should only do so in the presence of a 
qualified technician. 

 

4. General Description 

 
The DCM267 is a hand-held, battery-operated 3 ½ digital clamp meter for measuring DC and 
AC voltages, AC current, resistance, frequency, temperature (°C & °F), diodes and continuity. 
 

5. Description of the Front Panel 

 

Jaws 
They pick up the AC current flowing through the conductor. 

Hold-button 
Press this button to freeze the last reading and make the “H”-symbol appear on the display. 
Press this button again to make the “H”-symbol disappear. 

Function and Range Switch 
Used to select functions and desired ranges as well as to (de)activate the meter. 

Back light 
Press the (

) “LIGHT”-button to activate the back light for 

±

 3 to 5 seconds. 

Display 
3 ½ digit LCD with 7 segments. Height : 18mm. 

“VHz”-jack 
This is the positive (+) input terminal for tension and frequency measurements. Connect the 
red test lead to this input terminal. 

“COM”-jack 
This is the negative (-) input terminal for every type of measurement, with the exception of 
current measurements. Connect the black test lead to this input terminal. 

”-jack 

This is the positive input terminal for resistance, diode and continuity measurements. 
Connect the red test lead to this input terminal. 

Temperature measuring socket 
Connect a “K”-type thermocouple to this input terminal and contact the object to be measured 
with the thermocouple. 

Trigger 
Press the lever to open the jaws. The jaws will close again when the lever is released. 

Содержание DCM267

Страница 1: ...DCM267 DIGITAL CLAMP METER DIGITALE STROOMTANG PINCE AMPEREMETRIQUE DIGITALE STROMZANGE User Manual Gebruikershandleiding Manuel d utilisation Gebrauchsanleitung...

Страница 2: ...ments Just like all other electrical measuring instruments the DCM267 will respond to unwanted signals e g transients that may occur in the circuit being tested Exercise extreme caution when using thi...

Страница 3: ...current resistance frequency temperature C F diodes and continuity 5 Description of the Front Panel Jaws They pick up the AC current flowing through the conductor Hold button Press this button to free...

Страница 4: ...DCM267 GB 3...

Страница 5: ...jack and the black test lead to the COM jack The polarity of the red test lead is positive 2 Use the rotary switch to select the desired range 3 Connect the test leads to the resistor to be measured a...

Страница 6: ...h in the position 3 Connect the red test lead to the anode and connect the black test lead to the cathode of the diode to be tested 4 The approximate forward voltage drop of the diode will be displaye...

Страница 7: ...ce 10M 7 2 AC Voltage Range Resolution Accuracy 200V 0 1V 750V 1V 1 0 of rdg 3 digits Input impedance 10M Response average response calibration in rms of a sine wave Frequency range 40Hz 400Hz 7 3 AC...

Страница 8: ...e battery compartment Only use the device if the cover is in place and the screws are tightened in order to avoid electroshocks The symbol is displayed if the battery needs to be replaced Use the foll...