Multi-Color Low Battery Indicator

The Veinlite EMS PRO is powered by two AA alkaline batteries. The batteries will 

 

provide between 3 and 5 hours of continuous use depending on the setting and type 

of battery used. When the batteries run low, the green light under the button [2] will be 

replaced by an orange low battery indicator light. Approximately one hour of battery 

use is left when the orange low battery indicator is replaced by a red blinking light.  

Installing or Replacing Batteries

To install or replace the batteries, press firmly on the battery cap [3] and slide it off. 

Remove any used batteries. Install two new AA batteries, oriented as indicated by the 

markings inside the battery compartment then slide the battery cover back into place.

Attaching a Lanyard 

A lanyard can be attached to the Veinlite EMS PRO, via the hole at the bottom of the 

unit [4]. Use a thin cord for this purpose.  

3

2

Operation

• Place the Veinlite EMS PRO against the patient’s skin with the LED side in direct 

 

contact with the skin perpendicular to vein direction. Apply gentle pressure to 

 

maintain optimal skin contact.

• To turn on the LEDs, press the opaque button located on the left side of the device.
• Press the button a second time to switch on the white LEDs for surface illumination. 
• Press the button a third time to turn off.

Using the Disposable Plastic Cover

The single-use disposable plastic covers (Product #VEMS-DPC) for the Veinlite EMS 

PRO 

MUST be used when the devise is placed in contact with a patient’s skin. 

CAUTION: These plastic covers are not sterile. They should be wiped with a germicidal 

wipe prior to use.   

• Before using Veinlite EMS PRO, place it inside the disposable cover with the flap on 

the same side as the LEDs.

• Pull the flap over the device to the other side. 
• Wipe the LED side of the Veinlite EMS PRO and any part of it coming in contact with 

the patient, with a medically approved germicidal wipe. 

 

Precautions

    • Avoid using alcohol on the clear plastic window as this may cause fogging.

• Do not immerse the device in liquid.
• Do not autoclave the device. 

Scan Direction

a

b

Access Vein

To scan for veins, place device 

against skin 

perpendicular to vein  

direction

. Scan it forward and

 

backward. Veins will be visible as 

 

dark lines due to transillumination. 

Once a suitable vein has been  

identified, rotate device 90˚, 

 

parallel to vein. 

Apply gentle pressure on the 

 

Veinlite EMS PRO to secure the vein. 

Push back slightly on the Veinlite 

EMS PRO to stretch the skin and 

 

provide traction. 

Insert the needle into the vein, 

through the opening of the device.

 

NOTE: Discard the single-use disposable 

plastic cover after each patient. 

1

2

4

3a

3b

WARNING: DO NOT SHINE THE LED LIGHTS IN THE PATIENT’S EYES.

Attachments

Two patented*** attachments are included with the Veinlite EMS PRO: 

A. A snap-on light shield to reduce interference from bright overhead lighting.

B. A snap-on pediatric adapter to reduce the size of the viewing area for use with infants.

The diagrams below indicate how to attach the light shield and neonatal adapter. 

A. Veinlite EMS PRO Light Shield

B. Veinlite EMS PRO Pediatric Adapter