64
VEGAPULS 51K … 54K
21750-EN-031222
Supplement
1.4
Determination of safety-related characteristics
The failure limit values for a safety function, depending on the SIL class.
Safety integrity
Safety integrity
Safety integrity
Safety integrity
Safety integrity
Low demand mode
High demand or continuous
level
mode
SIL
PFD
avg
PFH
4
>10
-5
up to <10
-4
>10
-9
up to <10
-8
3
>10
-4
up to <10
-3
>10
-8
up to <10
-7
2
>10
-3
up to <10
-2
>10
-7
up to <10
-6
1
>10
-2
up to <10
-1
>10
-6
up to <10
-5
(from IEC 61508, part 1/7.6.2)
Safety integrity of the hardware:
Limitations due to the architecture for safety-related subsystems of type B
Safe failure fraction SFF
Hardware fault tolerance HFT
0
1 (0)
1)
2
< 60 %
not allowed
SIL 1
SIL 2
60 % up to < 90 %
SIL 1
SIL 2
SIL 3
90 % up to < 99 %
SIL 2
SIL 3
(SIL 4)
> 99 %
SIL 3
(SIL 4)
(SIL 4)
(from IEC 61508, part 2/7.4.3)
1)
Acc. to IEC 61511-1, paragraph 11.4.4 the fault tolerance specified acc. to the above chart can be reduced
by one for all subsystems if the following conditions are met:
- instrument has proven during operation
- only process-relevant parameters can be modified on the instrument (e.g. measuring range, …)
- the adjustment of this process-relevant parameters is protected (e.g. password, …)
- the safety function needs less than SIL 4
Содержание VEGAPULS 51K
Страница 1: ...Operating Instructions VEGAPULS 51K 54K 4 20 mA HART compact sensor...
Страница 41: ...VEGAPULS 51K 54K 41 21750 EN 031222 Rx VEGAMET VEGALOG Ri 250 250 Ri 250 Set up PLC PLC...
Страница 49: ...VEGAPULS 51K 54K 49 21750 EN 031222 Set up VEGALOG VEGAMET Rx...
Страница 70: ...70 VEGAPULS 51K 54K 21750 EN 031222 Supplement CE declaration of conformity...
Страница 71: ...VEGAPULS 51K 54K 71 21750 EN 031222 Supplement...