7 Diagnostics and servicing
VEGAPOINT 11 • Transistor (PNP)
56622-EN-221012
Out of specification
Red signal lamp flashes
Error
Cause
Rectification
S604
Overload on output
Switching output overloaded
Check electrical connection
Reduce switching load
Function check
Red signal lamp flashes
Error
Cause
Rectification
C700
Simulation active
Terminate simulation mode
7.4 How to proceed if a repair is necessary
You can find an instrument return form as well as detailed information
about the procedure in the download area of our homepage. By doing
this you help us carry out the repair quickly and without having to call
back for needed information.
Proceed as follows in case of repair:
•
Print and fill out one form per instrument
•
Clean the instrument and pack it damage-proof
•
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
•
Ask the agency serving you to get the address for the return ship-
ment. You can find the agency on our homepage.