76
11 Supplement
VEGAFLEX 83 • Profibus PA
44223-EN-140430
L
0
-2mm
(-0.079")
2mm
(0.079")
15mm
(0.591")
-15mm
(-0.591")
0,2 m
(7.874")
0,08 m
(3.15")
0,04 m
(1.575")
1
Fig. 49: Deviation VEGAFLEX 83 in rod version in water
1 Dead band - no measurement possible in this area
L Probe length
L
2mm
(0.079")
0
-2mm
(-0.079")
15mm
(0.591")
-15mm
(-0.591")
-10mm
(-0.394")
0,35 m
(13.78")
0,15 m
(5.906")
0,1 m
(3.937")
0,05 m
(1.969")
1
1
Fig. 50: Deviation VEGAFLEX 83 in rod version in oil
1 Dead band - no measurement possible in this area
L Probe length
Repeatability
≤ ±1 mm
Variables influencing measurement accuracy
Temperature drift - Digital output
±3 mm/10 K relating to the max. measuring range or
max. 10 mm (0.394 in)
Additional deviation through electromag-
netic interference acc. to EN 61326
< ±10 mm (< ±0.394 in)