background image

15

5 Set up with the display and adjustment module

VEGABAR 86 • Foundation Fieldbus

46323-EN-210430

Diagnosis:

 Information, for example, of device status, peak value, 

simulation

Additional adjustments:

 date/time, reset, copy function

Info:

 Instrument name, hardware and software version, factory cali-

bration date, device ID, sensor features

Note:

For optimum setting of the measuring point, the individual submenu 

items in the main menu item "

Setup

" should be selected one after 

the other and provided with the correct parameters. If possible, go 

through the items in the given sequence.
The procedure is described below.
The following submenu points are available:

The submenu points are described below.

5.4  Menu overview

The following tables show the adjustment menu of the instrument. 

Depending on the instrument version or application, all menu items 

may not be available or some may be differently assigned.

Setup

Menu item

Parameter

Default value

Application

Application

Level

Secondary sensor for electronic differen-

tial pressure

Deactivated

Units

Adjustment unit (m, bar, Pa, psi … user-

defined)

mbar (with nominal measuring range 

≤ 400 mbar)
bar (with nominal measuring ranges 

≥ 1 bar)

Temperature unit (°C, °F)

°C

Position correction

0.00 bar

Adjustment

Zero/Min. adjustment

0.00 bar
0.00 %

Span/Max. adjustment

Nominal measuring range in bar
100.00 %

Damping

Integration time

1 s

Linearisation

Linear, cylindrical tank, … user-defined

Linear

Lock adjustment

Blocked, released

Released

Содержание VEGABAR 86

Страница 1: ...Quick setup guide Submersible pressure transmitter with ceramic measuring cell VEGABAR 86 Foundation Fieldbus Document ID 46323...

Страница 2: ...ment module 11 5 1 Insert display and adjustment module 11 5 2 Parameter adjustment Quick setup 12 5 3 Parameter adjustment Extended adjustment 14 5 4 Menu overview 15 6 Set up with smartphone tablet...

Страница 3: ...s is a state of the art instrument complying with all prevailing regulations and directives The instrument must only be operated in a technically flawless and reliable condition The operator is respon...

Страница 4: ...equipment NE 53 Compatibility of field devices and display adjustment components NE 107 Self monitoring and diagnosis of field devices For further information see www namur de 1 7 Environmental instr...

Страница 5: ...ment With it you can find the following instrument data on our homepage Product code HTML Delivery date HTML Order specific instrument features HTML Operating instructions and quick setup guide at the...

Страница 6: ...or heated vessels Note Make sure that during installation or maintenance no moisture or dirt can get inside the instrument To maintain the housing protection make sure that the housing lid is closed...

Страница 7: ...7 3 Mounting VEGABAR 86 Foundation Fieldbus 46323 EN 210430 Instruments in protection IP66 IP68 1 bar ventilation via capillar ies in non detachable cable Instruments with absolute pressure...

Страница 8: ...n nut of the cable gland and remove blind plug 4 Remove approx 10 cm 4 in of the cable mantle strip approx 1 cm 0 4 in of insulation from the ends of the individual wires 5 Insert the cable into the s...

Страница 9: ...le chamber housing 1 Voltage supply signal output 2 Contact pins for the display and adjustment module or interface adapter 3 Simulation switch 1 mode for simulation release 4 For external display and...

Страница 10: ...s 5 1 2 6 7 8 2 3 4 1 Fig 6 Connection compartment double chamber housing 1 Voltage supply signal output 2 For display and adjustment module or interface adapter 3 For external display and adjustment...

Страница 11: ...d by 90 It is not necessary to interrupt the power supply Proceed as follows 1 Unscrew the housing lid 2 Place the display and adjustment module on the electronics in the desired position and turn it...

Страница 12: ...asily adapt the sensor to the application select the menu item Quick setup in the start graphic on the display and adjustment module Carry out the following steps in the below sequence You can find Ex...

Страница 13: ...the instrument offset on the measured value Max adjustment In this menu item you carry out the max adjustment for level Enter the percentage value and the corresponding value for the max level Min adj...

Страница 14: ...he real product level during the adjustment is not important because the min max adjustment is always carried out without changing the product level These settings can be made ahead of time without th...

Страница 15: ...nu points are available The submenu points are described below 5 4 Menu overview The following tables show the adjustment menu of the instrument Depending on the instrument version or application all...

Страница 16: ...ure Actual measuring cell and electronic tem perature Simulation Pressure percent current output linearized percent measuring cell tem perature electronics temperature Process pressure Additional adju...

Страница 17: ...te Bluetooth 1 Switch On Bluetooth active Off Bluetooth not active The security concept of Bluetooth operation absolutely requires that the default setting of the sensor PIN be changed This prevents u...

Страница 18: ...lect the device search via Bluetooth and start the search function The device auto matically searches for Bluetooth capable devices in the vicinity The message Instrument search running is displayed A...

Страница 19: ...19 6 Set up with smartphone tablet PC notebook via Bluetooth VEGABAR 86 Foundation Fieldbus 46323 EN 210430 Fig 11 Example of an app view Setup sensor adjustment App view...

Страница 20: ...ata version IP66 IP67 and IP66 IP68 0 2 bar 1 Options of the cable entry Cable entry M20 x 1 5 NPT Cable gland M20 x 1 5 NPT cable see below table Blind plug M20 x 1 5 NPT Closing cap NPT Material cab...

Страница 21: ...21 Notes VEGABAR 86 Foundation Fieldbus 46323 EN 210430...

Страница 22: ...22 Notes VEGABAR 86 Foundation Fieldbus 46323 EN 210430...

Страница 23: ...23 Notes VEGABAR 86 Foundation Fieldbus 46323 EN 210430...

Страница 24: ...ing scope of delivery application practical use and operat ing conditions of the sensors and processing systems correspond to the information available at the time of printing Subject to change withou...

Отзывы: