11
4 Mounting
Electronics module • SOLITRAC, MINITRAC, POINTRAC, WEIGHTRAC
40104-EN-210107
Keep in mind that the following work must only be carried out after an
appropriate training from VEGA.
Proceed as follows:
1. Switch off the source reliably - Set the source container to "Off"
Secure radiation source from being switched on again.
Warning:
Always switch off the radiation source when working on the sensor
and secure it against unauthorized or unintentional switching on.
2. Switch off voltage supply
Warning:
Connect only in a voltage-free state. Switch off the voltage supply reli-
ably and secure it against unauthorized or unintentional switching on.
3. Carry out all work in a clean and dust-free environment. If pos-
sible, dismount the sensor.
Caution:
The black rubber protective cap protects the light-sensitive photomul-
tiplier against incident light. Only remove the protective cap when you
are requested to do so in this manual.
4. Check if all parts are included:
- Optical pad
- Plastic spatula
- Special grease (tube)
- Coded pins on mounting support (2 pcs.)
5. Unscrew cover (1) of the electronics compartment (5)
6. Loosen and detach the terminal blocks (6) of the old electronics
module on the lateral "clamping brackets" (terminal blocks not
included in the delivery)
Содержание PT30E.MP
Страница 1: ...Operating Instructions Electronics module SOLITRAC MINITRAC POINTRAC WEIGHTRAC Document ID 40104...
Страница 25: ...25 Notes Electronics module SOLITRAC MINITRAC POINTRAC WEIGHTRAC 40104 EN 210107...
Страница 26: ...26 Notes Electronics module SOLITRAC MINITRAC POINTRAC WEIGHTRAC 40104 EN 210107...
Страница 27: ...27 Notes Electronics module SOLITRAC MINITRAC POINTRAC WEIGHTRAC 40104 EN 210107...