80
11 Supplement
VEGAPULS 69 • Two-wire 4 … 20 mA/HART
47247-EN-200427
Electromechanical data - version IP66/IP68 (1 bar)
Options of the cable entry
Ʋ Cable gland with integrated connec-
tion cable
M20 x 1.5 (cable ø 5 … 9 mm)
Ʋ Cable entry
½ NPT
Ʋ Blind plug
M20 x 1.5; ½ NPT
Connection cable
Ʋ Wire cross-section
0.5 mm² (AWG 20)
Ʋ Wire resistance
< 0.036 Ω/m
Ʋ Tensile strength
< 1200 N (270 lbf)
Ʋ Standard length
5 m (16.4 ft)
Ʋ Max. length
180 m (590.6 ft)
Ʋ Min. bending radius (at 25 °C/77 °F)
25 mm (0.984 in)
Ʋ Diameter
approx. 8 mm (0.315 in)
Ʋ Colour - Non-Ex version
Black
Ʋ Colour - Ex-version
Blue
Interface to the external display and adjustment unit
Data transmission
Digital (I²C-Bus)
Connection cable
Four-wire
Sensor version
Configuration, connection cable
Cable length
Standard cable
Special cable
Shielded
4 … 20 mA/HART
50 m
●
–
–
Profibus PA, Founda-
tion Fieldbus
25 m
–
●
●
Integrated clock
Date format
Day.Month.Year
Time format
12 h/24 h
Time zone, factory setting
CET
Max. rate deviation
10.5 min/year
Additional output parameter - Electronics temperature
Range
-40 … +85 °C (-40 … +185 °F)
Resolution
< 0.1 K
Deviation
±3 K
Output of the temperature values
Ʋ Indication
Via the display and adjustment module
Ʋ Output
Via the respective output signal
Voltage supply
Operating voltage U
B
12 … 35 V DC