18.1.4 Fuse Characteristics and Replacement
•
NanoScope Controller
: 6 fuses; 3 identical fuses in use and 3 spare fuses. The fuse
characteristics are silk-screened on the back of the NanoScope controller (see
Table 18.1a
NanoScope Controller Fuse Characteristics
•
Illuminator Circuit board
: 2 fuses; one fuse is non-replaceable (281-000-024), while
the other fuse is replaceable and is shipped with a spare (see
•
Vacuum Pump Board
: 2 fuses; each has a spare (see
Table 18.1b
Illuminator Circuit Board and Vacuum Pump Board Fuse Characteristics
Fuse Replacement Procedure
1. Remove all power from the system.
2. Remove the lid of the appropriate control box (4 screws).
Line
Voltage
Fuse Selection
F1
F2
F3
100V
2.0A
800mA
800mA
120V
2.0A
800mA
800mA
220V
1.0A
400mA
400mA
240V
1.0
400mA
400mA
All Fuses: 250V, Time-lag, 5mm x 20mm
Illuminator
Board
Part Number
Current
Dimensions
Max
Voltage
Time
Delay
Other
100V
281-000-026
1.25A
5 x 20mm
250V
Yes
IEC
120V
281-000-032
1.00A
5 x 20mm
250V
Yes
IEC
240V
281-000-031
0.5A
5 x 20mm
250V
Yes
IEC
Vacuum Pump
Board
100V
281-000-030
0.25A
5 x 20mm
250V
Yes
IEC
120V
281-000-029
0.2A
5 x 20mm
250V
Yes
IEC
240V
281-000-028
0.125A
5 x 20mm
250V
Yes
IEC
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......