Calibration
Check Sample Orthogonality
Rev. D
Dimension 3100 Manual
315
Note:
In
, pits align with the vertical (slow) axis but skew with the
horizontal (fast) axis. The angle should be measured with the vertex near the
center of the image and the vertices in the upper-right or lower-left quadrant.
The angle displays on the white status bar at the bottom of the display monitor.
3. Use the mouse to drag the cursor until it is oriented correctly then read the angle off the status
bar. If the angle differs by more than a half degree, a correction is required; otherwise, move
on to the next step.
17.3.2 Adjust Sample Orthogonality
1. Click
Realtime
>
Microscope
>
Calibrate
>
Scanner
to access the
Scanner Calibration
panel. The
Orthogonality
parameter is displayed on the bottom-right corner of the panel.
2. Enter the difference between 90º and the angle measured in the
Top View
image.
Note:
For example, if the angle measured in the
Top View
image was 92.5º, enter a
value of
-2.5
º in the
Orthogonality
parameter.
3. Click
OK
to exit the
Scanner Calibration
panel
4. Capture another image and re-measure the angle.
5. Repeat correction of
Orthogonality
until the scanned image shows less than 0.5º of error.
Note:
After a major change to the orthogonality parameter, you may need to
physically realign the calibration standard to the image frame.
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......