Calibration
Check Sample Orthogonality
314
Dimension 3100 Manual
Rev. D
17.2.4 Set Up Contact AFM
1. Set the
AFM mode
to
Contact
.
Note:
The microscope can be calibrated using STM; however, this example utilizes
contact AFM.
2. Set the
Scan angle
to
0
degrees.
3. Adjust Realtime parameters to obtain a good-quality, maximum
Scan size
image (
440V
). Set
the
Scan rate
to
2.44 Hz
and the
Number of samples
parameter to
256
.
4. With the sample engaged, check the scanning line relative to the reference’s features; the
scan should be orthogonal to the pits on the reference.
5. If the reference requires rotation,
Withdraw
and rotate the sample to improve orthogonality
between sample and scan line.
6. Repeat until features are oriented orthogonally with the scan frame.
17.3 Check Sample Orthogonality
Check the sample scan for orthogonality along both the X- and Y-axes. If the scan is aligned along
one axis of the scan but not another, it may be necessary to adjust the microscope’s
Orthogonality
parameter in the
Scanner Calibration
panel.
17.3.1 Measure Orthogonality
1. To measure a captured image’s orthogonality, view it using the
Offline
>
View
>
Top View
function.
2. On the display monitor, select the
Angle
command and use the mouse to draw a cursor
between the edges (or centers) of widely spaced pits (see
Figure 17.3a
Non-Orthogonal and Corrected, Orthogonal Image
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......