Magnetic Force Microscopy
Software Setup Configuration (Phase, Quadrex or NSIV)
Rev. D
Dimension 3100 Manual
269
Important Points
1. Extender-compatible microscope electronics are required to permit operation of the phase
detection extender option. Standard electronics on these microscopes require hardware
upgrades. Consult Veeco for details.
2. Turn off the power to the NanoScope controller whenever connecting or disconnecting the
Extender.
15.4.2 Quadrex Extender
The Quadrex Extender connects to a NanoScope IIIa in the same way as the Phase Extender
described above, and has similar hardware requirements. Quadrex operation requires NanoScope
software version 5.12 or higher.
15.4.3 NanoScope IV
NanoScope IV has the Quadrex electronics integrated in to the design and requires no modification
or additions for MFM.
15.5 Software Setup Configuration (Phase, Quadrex or
NSIV)
1. Select
di
>
Microscope Select
to display the
Microscope Select
dialog box (see
Figure 15.5a
Microscope Select Dialog Box
2. Click the
Edit
button to open the
Equipment
dialog box.
3. Select the appropriate
Controller
.
4. Select the appropriate
Extender
.
012