Magnetic Force Microscopy
Basic MFM Operation
262
Dimension 3100 Manual
Rev. D
8. Under
Scan Controls
, set the
Scan size
to
5µm
and
Scan rate
to
1–2Hz
.
9. Set the
Channel 1
Data type
to
Height
, the
Data scale
to
75nm
and the
Line direction
to
Retrace
.
10. Engage the AFM and make the necessary adjustments to obtain a good topographical image
while displaying height data.
Figure 15.2e
Topographic (left) and Magnetic Force Gradient Image (right)
Note:
The MFM data displays in Channel 2; however, the parameter settings are
different depending on whether Phase Detection or Amplitude Detection is
being used
CAUTION:
Use the highest possible
Setpoint
to ensure that the tip is
contacting the surface only lightly. The image should be similar to
the topographic image shown on the left of
. The
surface is fairly flat with lubrication nodules of various sizes. A
good image of the nodules indicates that the tip is sharp.