Force Calibration Control Panels and Menus
212
Dimension 3100 Manual
Rev. D
13.3 Force Calibration Control Panels and Menus
The
Force
Calibration
Control
window (see
) manipulates the microscope in
Force
Calibration
mode. The parameters control the rate, start position and amplitude of the triangle
wave applied to the Z piezo. You can also adjust the
Setpoint
value of the cantilever deflection
voltage used in the feedback loop during imaging. The
Capture
button stores the force curve for
Offline
viewing. Some parameters influence the operation of the microscope during imaging; most
of the menu parameters affect only
Force Calibration
mode.
Figure 13.3a
Advanced Force Calibration Control Window (Contact Mode AFM)
There are 2 methods to enter Force Calibration Mode:
•
View
>
Force Mode
>
Calibrate
will open a simplified subset of the parameters and
commands available.
•
View
>
Force Mode
>
Advanced
includes the full set of available parameters and
commands.
Additionally:
•
View
>
Force Mode
>
Step
enters force mode with the parameter
start mode
=
step
and the piezo fully retracted.
•
View
>
Force Mode
>
Volume
enters Force Volume Mode.
Items in the
Force Calibration
Control
Window are discussed individually below.
038
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......