Scanning Tunneling Microscopy (STM)
Etching Tungsten Tips
192
Dimension 3100 Manual
Rev. D
11.6 Etching Tungsten Tips
You can purchase tungsten tips from Veeco or make them yourself. This section describes the
process of etching tungsten tips for use with the NanoScope.
Materials Required
•
Variac auto transformer
•
Optical microscope (20-100X)
•
Sodium nitrite (NaNO2)
•
Distilled water
•
Ethyl alcohol
•
WD-40 (anti-oxidant)
•
Two 50ml beakers
•
Tipholder
•
Platinum wire
•
Tungsten Wire, 0.010" Diameter
•
Miscellaneous Wire/Clips
11.6.1 Procedure
1. Mix a 5 percent (by weight) solution of Sodium Nitrite in water.
2. Pour
≈
40ml of the Sodium Nitrite solution into a beaker.
3. Pour
≈
40ml of WD 40 into a beaker.
4. Construct an electrode out of the platinum wire and insert it into the beaker.
5. Adjust the variac for 30V, and with it Off, connect one output to the platinum electrode.
6. Cut 10 to 12 pieces of tungsten wire
≈
1.25cm long. Before etching, check to make sure that
at least one end of the wire has not split by inserting the ends into a tip holder. If an end has
split, you will not be able to insert it into the tipholder. You can etch the end that has split,
however, preserve the unspoiled end.