Contact AFM
Basic Contact Mode AFM Operation
124
Dimension 3100 Manual
Rev. D
Feedback Controls Panel
1. Set the
Integral gain
to
2.0
and the
Proportional gain
to
4.0
2. Set the
Deflection Setpoint
to
0V
.
Figure 8.1e
Suggested Feedback Controls Settings
Channel Panels 1 and 2
1. In the
Channel 1
panel, set
Data type
to
Height
(see
2. Set
Data scale
to a reasonable value for the sample.
Note:
For example, for a 200nm step height calibration sample, a reasonable Data
scale setting is 300nm initially.
3. Set
Line direction
to either
Trace
or
Retrace
.
4. On the
Channel 2 panel,
verify
Data type
is set to
Off
to disable the panel.
Figure 8.1f
Suggested Channel Controls Settings
018