Electric Force (EFM) Imaging
Surface Potential Detection—Preparation
Rev. B
MultiMode SPM Instruction Manual
263
14.6 Surface Potential Detection—Preparation
It is often desirable to apply a voltage to one or more areas of a sample. This may be done in two
ways: by connecting a voltage to the sample mounting chuck, or by making direct contact to the
sample. In both cases, jumper con
fi
gurations in the bottom of the microscope must be changed to
match the environment desired.
Note:
In addition to any recon
fi
gured jumpers, remember to connect the common or
negative terminal of an external power supply to the MultiMode ground or the
AFM chassis.
1. Power down the NanoScope controller and turn off all peripherals. Unplug the NanoScope
power cable from the microscope’s electronic box.
2. Locate the jumpers on the electronic board, visible in a “window” on the bottom of the
. As shipped from the factory, the baseplate jumpers should
appear as shown in
.
Figure 14.6a
Normal jumper con
fi
guration as shipped from factory
(for systems
with
the Basic Extender Module). Sample is held at ground.
3. Depending upon whether voltage is to be applied to the sample directly or indirectly,
reconfigure jumpers if indicated.
Ground
GND/OSC + DC
Unused
Auxiliary D (to NanoScope III controller)
To AFM Tip
Indicates jumpers
GND/OSC + DC
STM
Piezo Cap