Setup

Basic Assembly and Eyepiece Focusing

Ensure that the work area is free of dust and dirt.

1.  Remove all parts from packaging and place the polarimeter body on a stable 

work surface.

2.  Insert female end of the main power cord into the power jack located on the rear 

of the polarimeter.
3.  Insert male end of the main power cord into a suitable power outlet.
4.  Flip the power switch to the on position, labeled “|”.

5.  The C25L polarimeter utilizes an LED illuminator and therefore requires no warm up period before use.
6.  Observe the field of view through the eyepiece and rotate the eyepiece focus adjustment (figure 2) until the 

edges of the round field of view are sharply focused.

Operation

Glass Sample Cells

The polarimeter comes with 100mm and 200mm glass sample cells. Both utilize bubble traps to prevent air 

bubbles from giving inaccurate readings.

1.  Obtain a well homogenized sample of the substance to be tested.
2. Unthread and remove the endcap pieces from one end of the sample cell.
3.  Pour the substance into the sample cell leaving only a very small amount of air in the cell. Replace the endcap 

being careful to tighten only by hand and only until it prevents liquid from leaking out. Over tightening the end cap 

will cause stress in the glass which can produce erroneous readings.

4.  Open the sample chamber trap door and place the filled sample cell into the chamber with the bubble trap toward 

the top, making sure that any air bubbles still in the cell are trapped inside the bubble trap. Close the trap door.

Measuring

This instrument works by passing linear polarized light through an optically active 

liquid sample and measuring the degree of rotation that occurs. The reading is then 

displayed to the user through a triplex field of view with the left half-shadow and the 

right half-shadow being divided by a center section (figure 3).

Figure 2

Figure 6

Figure 4

Figure 3

Figure 5

1.  While looking through the eyepiece rotate the Vernier scale adjustment knob (figure 4) until the two half-shadows 

are at the same dark intensity as the center section (figure 5).
2.  The angle of rotation can now be read on the Vernier scales (figure 6) through the magnifying lenses located 

on the eyepiece. A reading of 0° indicates that the sample is optically inactive.

Содержание C25L

Страница 1: ...Operation Manual Model C25L Cat No 4601P Measurement Range 180 Manual Polarimeter...

Страница 2: ...eliable at the time of printing However no responsibility shall be assumed by VEE GEE Scientific for its use The information contained in this document is subject to change without notification This p...

Страница 3: ...umentforservice Safety Precautions Maintenance Fuse Replacement This instrument uses a 1A 250V part number 1200 FS4 fuse which is located on the bottom of the instrument 1 Turn the power switch off la...

Страница 4: ...are functioning and focusing properly and that the following items are supplied with your instrument 1 Assembled Polarimeter Body 1 ea 2 Power Cable NEMA 5 or Euro Schuko depending on voltage 1 ea 3 1...

Страница 5: ...for more information Item Part No Factory replacement 100mm Sample Cell with bubble trap 4600P SC100 Factory replacement 200mm Sample Cell with bubble trap 4600P SC200 1A 250V Fuse 1200 FS4 Replacemen...

Страница 6: ...nto the sample cell leaving only a very small amount of air in the cell Replace the endcap being careful to tighten only by hand and only until it prevents liquid from leaking out Over tightening the...

Страница 7: ...ns Operation continued Figure 7 Instrument Accuracy Your VEE GEE polarimeter was calibrated at the factory during production Follow the instructions below to check the accuracy of your instrument and...

Страница 8: ...Printed in USA No VGMNL092319 C25L...

Отзывы: