Range:
Brix 0.0-32.0%
Resolution:
0.2%
Accuracy:
±0.2%
Dimensions:
40 x 40 x 165mm (1.6 x 1.6 x 6.5”)
Weight:
240g (8.5 oz.)
Supplied With:
Vinyl Carrying Case (1), Plastic Transfer Pipet (1)
Printed in
China
VGMNL
010320
-43001
Operation Manual
Model BX-1
Brix 0-32%
Specifications
Cat. No. 43001
Temperature Correction Table
For Refractometers Calibrated At 20°C (
l
=
589nm)
Source:
ICUMSA, 1974
As an alternative to recalibration of the refractometer due to ambient temperature changes, the following tempera-
ture correction table can be used. Ensuring that both the distilled water and the ambient temperature are both at
exactly 20°C, follow steps 2-7 under the CALIBRATION section on the preceding page. Once this is accomplished, if
ambient temperature changes, simply apply the temperature correction values listed below. For example, a reading
of Brix 40.0% at 25°C is corrected to Brix 40.4%; a reading of Brix 65.0% at 10°C is corrected to Brix 64.25%.
Warranty information and
registration form can be found at:
www.veegee.com/service_support