Vector Lab Products Refrigerators and Freezers
17
Door Gasket Replacement
Fig. 5
1. Remove existing gasket from mounting track (Fig. 5) .
2. Verify mounting track is free of any remaining gasket material .
3. Align new gasket with mounting track and press firmly in place .
4. Open and close door, checking for proper gasket seal without
pinching against refrigerator .
Modular Refrigeration System
Vector Lab Products’ Modular Refrigeration System allows technicians to
remove the entire unit from the refrigerator top .
1. Remove rear support column using 5/16" nut driver .
2. Remove incoming power supply assembly .
3. Remove bolts securing unit cooler to refrigerator top .
4. Gently lift off cooler .