11 Appendix
35 / 41
value specifies the slope in counts per °C and is represented
in 10.6 bits fixed point two's complement notation. See
parameter m=<p8> for more details about drift compensation.
Changing this parameter may void ex factory calibration.
o=<p16> : Offset o of the linear drift compensation term n*T + o. The
value specifies an offset in counts and is represented
in 16 bits fixed point two's complement notation. See
parameter m=<p8> for more details about drift compensation.
Changing this parameter may void ex factory calibration.
p=<p16> : User specified parameter for the temperature drift
compensation. This is a value represented in 16 bits fixed
point two's complement notation. The value is an additional
offset value to fine adjust the camera. See parameter m=<p8>
for more details about drift compensation.
s=<p8> : Configure the UART(s) for main serial communications. Changes
are activated immediately.
<eaa*_bbbb>
|\| \__|
| | +---b: Baud rate of the main UART and (if
| | availiable) the second channel UART. The
| | second channel UART supports 9600 to 115200
| | Baud only.
| | 0 -> 110 Baud.
| | 1 -> 300 Baud.
| | 2 -> 600 Baud.
| | 3 -> 1200 Baud.
| | 4 -> 2400 Baud.
| | 5 -> 4800 Baud.
| | 6 -> 9600 Baud.
| | 7 -> 19200 Baud.
| | 8 -> 38400 Baud.
| | 9 -> 57600 Baud.
| | A -> 115200 Baud.
| +---------a: Second channel UART configuration. This
| channel is equivalent to the main UART
| concerning the use as command interface.
| But it is not intended to use both UARTs at
| the same time, because they share the same
| receive buffer. Output characters are always
| sent to both channels.
| 00 -> Second channel UART off. The active
| serial communication port is selected
| by hardware jumper settings.
| 01 -> Activate second channel via the frame
| grabber's serial port. This is either
| Camera Link TG/TC or Pleora device
| port serial 0.
| 10 -> Activate second channel via Pleora
| device port bulk 0 (MODE:UART,
| GE interface only).
| 11 -> Reserved.
+-----------e: Echo suppression.
0 -> Each character received from the host
is echoed back to it.
1 -> No echo.
v=<p8> : View file. Dumps all bytes of the specified file to the
serial interface. Raw data is output. If the file contains
binary data it may disturb a connected terminal program.
Please consider the available baud rate: large files may take
several minutes to transfer.
x=1 : Dump the complete processor RAM contents in hexadecimal
format. (*)
Command Overview (Upper Case Commands, Mainly for User Configuration)
---------------------------------------------------------------------
A=<p8> : File number of the first cold (low) reference image
of the two point correction (Gain Offset Correction).
Changing this parameter may void ex factory calibration.
0 : No loading of any correction image. The
memory content remains uninitialized at
VDS Vosskühler GmbH
IRC-320