Appendix
9. Code 8
●
Message : Bad Pixel's map correction error.
●
Solution : Restore the EzSensor’s calibration data from the installation CD or re-calibrate the sensor.
Contact Customer Service.
10. Code 9
●
Message: Bad Pixel's map correction error.
●
Solution: Restore the EzSensor’s calibration data from the installation CD or re-calibrate the sensor.
Contact Customer Service.
11. Code 10
●
Message :. Can't find dark frame.
●
Solution : Restore the EzSensor’s calibration data from the installation CD or re-calibrate the sensor.
Contact Customer Service
12. Code 11
●
Message: Can't find bright frames for calibration
●
Solution: User canceled image capture. Please try again.
13. Code 12
●
Message: Canceled image capturing.
●
Solution : User canceled image capture.
Please try again.
14. Code 13
●
Message: Wrong image processing parameters.
●
Solution: Check X-ray source.
15. Code 14
●
Message: Error in configuration file.
●
Solution: Please re-instal acquisition software.
16. Code 15
●
Message: Image process library was damaged.
●
Solution: Please re-instal acquisition software.