A.4
Electromagnetic Compatibility Information
Guidance and manufacturer’s declaration – electromagnetic emissions
The Product is intended for use in an electromagnetic environment as specified below. The customer or
the user of the Product should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Product uses RF energy only for its internal functions.
Therefore, its RF emissions are very low and are not likely to
cause any interference to nearby electronic equipment.
RF emissions
CISPR 11
Class A
The Product is suitable for use in all establishments, including
domestic establishments, and those directly connected to a
personal computer USB port used for domestic purposes.
Harmonics
emission
IEC 61000-3-2
A
Voltage
fluctuation
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic emissions
The Product is intended for use in an electromagnetic environment as specified below. The customer or
the user of the Product should ensure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
6 kV contact
±
8 kV air
±
6 kV contact
±
8 kV air
Floors should be wood, concrete or
ceramic tiles. If floors are covered with
synthetic material, relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±
2 kV for power
supply lines
±
1 kV for
input/output lines
±
2 kV for power
supply lines
±
1 kV for
input/output lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±
1 kV differential
mode
±
2 kV common
mode
±
1 kV differential
mode
±
2 kV common
mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips,
short interruption,
and voltage
variations on
power supply
input lines
IEC 60601-4-11
< 5 % U
T
(> 95 % dip in U
T
)
for 0.5 cycles
40 % U
T
(60 % dip in U
T
)
for 6 cycles
70 % U
T
(30 % dip in U
T
)
for 30 cycles
< 5 % U
T
(> 95 % dip in U
T
)
for 5 s
< 5 % U
T
(> 95 % dip in U
T
)
for 0.5 cycles
40 % U
T
(60 % dip in U
T
)
for 6 cycles
70 % U
T
(30 % dip in U
T
)
for 30 cycles
< 5 % U
T
(> 95 % dip in U
T
)
for 5 s
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the Product
requires continued operation during
power mains interruptions, it is
recommended that the Product be
powered from an uninterruptible power
source or battery
.
Power frequency
3.0 A/m
3.0 A/m
Power frequency magnetic fields should