Appendix
EzRay Chair (VEX-S300C) User Manual
69
EN
G
LISH
Guidance and manufacturer’s declaration - electromagnetic immunity
The EzRay Chair (VEX-S300C) is intended for use in the electromagnetic
environment specified below. The customer or the user of the EzRay Chair (VEX-
S300C) should assure that it is used in such an environment.
Immunity test
IEC 60601 Test
level
Compliance
level
Electromagnetic
environment -guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV Contact
±8 kV air
±6 kV Contact
±8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output
lines
±2 kV for power
supply lines
±1 kV for
input/output lines
Main power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Main power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
< 5 % U
т (>
95 % dip in U
т)
for 0.5cycle
40 % U
т (60 %
dip in U
т
) for 5
cycle, 6 cycle
70 % U
т (30 %
dip in U
т)
for 25
cycle, 30 cycle
<5 % U
т (<
95 % dip in U
т)
for 5 s
< 5 % U
т (> 95 %
dip in U
т)
for
0.5cycle
40 % U
т (60 %
dip in U
т)
for 5
cycle, 6 cycle
70 % U
т (30 %
dip in U
т)
for 25
cycle, 30 cycle
<5 % U
т (< 95 %
dip in U
т)
for 5 s
Main power quality
should be that of a
typical commercial or
hospital environment. If
the user of the EzRay
Chair (VEX-S300C)
image intensifier is to
continue to operate
during mains
interruption, it is
recommended that the
EzRay Chair (VEX-
S300C) be powered
from the uninterruptible
power supply.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE: U
т
is the V a.c. mains voltage before application of the test level.
Содержание EzRay Chair
Страница 1: ......
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...Notice ii EzRay Chair VEX S300C User Manual This page intentionally left blank...
Страница 16: ...1 General and Regulatory Information 8 EzRay Chair VEX S300C User Manual This page intentionally left blank...
Страница 20: ...2 Safety Instructions 12 EzRay Chair VEX S300C User Manual This page intentionally left blank...
Страница 44: ...4 Operation 36 EzRay Chair VEX S300C User Manual This page intentionally left blank...
Страница 48: ...6 User Maintenance 40 EzRay Chair VEX S300C User Manual This page intentionally left blank...
Страница 52: ...8 Disposing of the Unit 44 EzRay Chair VEX S300C User Manual This page intentionally left blank...
Страница 54: ...9 Product Specifications 46 EzRay Chair VEX S300C User Manual...
Страница 55: ...9 Product Specifications EzRay Chair VEX S300C User Manual 47 ENGLISH...
Страница 56: ...9 Product Specifications 48 EzRay Chair VEX S300C User Manual...
Страница 60: ...9 Product Specifications 52 EzRay Chair VEX S300C User Manual 4 Tube Dimensions mm...
Страница 80: ...0 Appendix 72 EzRay Chair VEX S300C User Manual This page intentionally left blank...
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......
Страница 86: ......