3
. Installing the Equipment
EzRay Air W (Model: VEX-S300W) Installation Manual
27
EN
G
LISH
3.3
Installing the Horizontal Arm
The length of the horizontal arm:
L= 450, 600, and 900 mm
Depending on arm’s length, the height where control unit is to be
installed is of critical importance.
The installation procedures are the same regardless of the arm’s
length.
1. After pulling out the cables through the opening, insert the horizontal arm into the hole
on the Power Box Assembly.
2. Fix the block assembly using the anti-rotation block2 (
Part No. 2
), anti-rotation block
stiffner (
Part No. 3
) and wrench bolts (
Part No. 12
).
Allen wrench
3 mm
Horizontal arm
Cable included
ANTI ROTATION
BLOCK2
Part No.:16
Qty.: 1 ea
ANTI ROTATION
BLOCK
STIFFNER
Part No.: 17
Qty.: 1 ea
WRENCH BOLT
Part No.: 20,
w/spring and flat
washer
Size: M5 x 20
Qty.: 2 ea
Содержание EzRay Air W
Страница 1: ......
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 10: ...Table of Contents x EzRay Air W VEX S300W Installation Manual This page intentionally left blank...
Страница 12: ...1 Introduction 2 EzRay Air W Model VEX S300W Installation Manual This page intentionally left blank...
Страница 54: ...4 Product Specifications 44 EzRay Air W Model VEX S300W Installation Manual...
Страница 60: ...4 Product Specifications 50 EzRay Air W Model VEX S300W Installation Manual This page intentionally left blank...
Страница 61: ...Appendix EzRay Air W Model VEX S300W Installation Manual 51 ENGLISH Appendix Drilling Template...
Страница 64: ...Appendix 54 EzRay Air W Model VEX S300W Installation Manual This page intentionally left blank...
Страница 66: ......
Страница 67: ......
Страница 68: ......