Vatech A9 (PHT-30CSS)
Technical manual
VDH-TeM-098
140 / 214
COPYRIGHT© 2020 VATECH
.
ALL RIGHTS RESERVED
4.
Homogeneity evaluation
4.1. Homogeneity standards
Parameter
ROI 1
ROI 2
ROI 3
Remark
Homogeneity
(Ave)
8000 or more
1200 or more
The difference between ROI 1,2
values is 10% or less
4.2. Phantom Specification
QUART Phantom
QUART - Ball Phantom
4.3. Test Method
The evaluation is conducted in the same manner as in the evaluation method in Section 3.3.
4.4. Test Result
Parameter
ROI 1 (600, 800) ROI 2 (2550, 800) ROI 3 (1574, 800)
Pass/Fail
Homogeneity(Ave)