background image

Product Name:

 

Oxygen Generator

Model number:

 

NT-3

Scope  of  application:

  It  is  only  used  for  health  preservation  and  improving  the 

quality of life, not for medical treatment or life continuation.

Oxygen production method:

 pressure swing adsorption separation method (PSA 

method)

Work rate:

 30W

Input voltage:

 DC12V~16. 8V

Four levels of timing:

 40 minutes, 60 minutes, 80 minutes, 180 minutes

Product size: 

20

.

5 * 9

.

5 * 13cm

Product net weight: 

1.4Kg

Maximum oxygen production flow (equivalent):

 3L

/

min

Oxygen concentration: 

32

.

5%±2

.

5

(v/v)

Standard battery pack capacity: 

2500mAH 14.8V

the highest voltage is 16.8V

The longest use time: 

2

.

5h.

Change a new power
or battery

1

Change a new power
or battery

2

Change a new power
or battery

3

Change a new power 
board

4

Not working

Not  connected
to power supply

Power broken

The charger 
doesn’t work

Power line is broken

No AC

Contact the agent or 
manufacturer to deal 
with it.

5

Internal damage

Not working 

Internal  failure 
of the machine

Change the correct 
power supply

6

Incorrect or damaged
power supply

The  first  light 
flashes slowly

The power 
supply is 
abnormal

Change the correct 
power supply

7

Incorrect or damaged
power supply

The first light 
flashes quickly

The power 
supply is 
abnormal

Smooth  the  air  inlet 
and  outlet  of  the 
machine and wait for
the  temperature  of 
the  machine  to  drop 
to normal.

8

Poor heat dissipation

The second light 
flashes slowly

High internal 
temperature

Put  the  machine  in 
a n   e n v i r o n m e n t  
h i g h e r   t h a n   5  
degrees Celsius for
4   h o u r s   b e f o r e  
using it.

9

Working environment
temperature is too low

The second light 
flashes quickly

The ambient 
temperature 
is too low

No

10

No

There is a rhythm 
like breathing

Normal working 
sound

Poor heat dissipation, 
the  machine  can  still 
work,  this  time  will 
increase  the  heat 
dissipation capacity.

11

Poor heat dissipation

Fan noise is
too loud

Temperature 
is too high 

Product specifications

Troubleshooting for general problems

NO.

Phenomenon

Problems

Reason

Solution

- 8 -

- 9 -

Содержание NT-3

Страница 1: ...Oxygen Generator User Manual NT 3 Agreen life with oxygen begins now...

Страница 2: ...generator and keep this instruction for future reference or forwarding to any future users This manual explains and stipulates the operating conditions and performance specifications operating environ...

Страница 3: ...lly charge is about 4 10 hours The use of oxygen generator 1 Please place the product correctly ensure that the air inlet and outlet are unobstructed and the distance from the outside to other objects...

Страница 4: ...e 2 Standard battery pack 3 Take out the nasal suction tube and insert the air inlet of the nasal suction tube into the air outlet of the oxygen generator 4 Put on the nasal straw as shown in the pict...

Страница 5: ...g the oxygen inhaler be careful not to bend the pipeline to prevent the air path from being blocked 2 Properly keep and place the battery pack to prevent accidents and help prolong the life of the pro...

Страница 6: ...o deal with it 5 Internal damage Not working Internal failure of the machine Change the correct power supply 6 Incorrect or damaged power supply The first light flashes slowly The power supply is abno...

Страница 7: ...consent of the manufacturer 6 Products damaged due to circumstances beyond the control of the manufacturer such as earthquakes floods etc 7 Products repaired by anyone other than the manufacturer s se...

Страница 8: ...12 13...