1. Turn on the scale by pressing the [ON] key.
2.
Wait for the scale to go to zero first, then press and hold the [ON] key until the display
show “0”, then release it. Then the required calibration weight will be displaying and
blinking. (if you do not have right calibration weights, you can press the button “TARE”
to select a weight).
3. At this point, place the required calibration weight on the scale. After a few minutes,
stop blinking, backlight is brighter, then the weight will display again.
4. Calibration is complete!
4. CALIBRATION
(only perform calibration if there is a problem)
•
Take off the scale cover carefully. Put the scale on a stable platform, without shaking
and wind.
•
Do not impact the scale try and operate the scale carefully.
•
Keep the scale out of your mobile over 50 cm.
•
When “S” is display on left up, you can read the weight data.
•
When “Lo” is display, please replace the batteries.
•
Please calibrate the scale when temperature changed too much. Place the scale for
30 minutes first, let the temperature is even, then make the calibration.
•
Do not overload to avoid to damage the loadcell. Once overload, “FULL” is displayed.
5. NOTE