VHS-4, VHS-6 and VHS-250 Diffusion Pumps
D
R
A
FT 3/9
/
05
4-3
Inlet pressure surges
Incorrect heater voltage.
Check and correct voltage.
Fluid outgassing.
Condition pump fluid by
operating pump for a few hours.
Leak in system ahead of
pump inlet.
Check and correct.
High chamber contamination
Forepressure too high.
Check for leak in foreline, poor
mechanical pump performance,
breakdown of pump fluid, and
incorrect valve operation.
Prolonged operation at high
throughput at pressure above
10
−
3
Torr (1.33x10
-3
mbar).
Review operating procedures.
Improper system operation
air-release procedures.
Review operating procedures
Pump will not start
Safety circuits and/or protective
devices prevent contactor from
staying closed.
Check utilities, flow switches,
interlocks, and thermostat
operation.
Table 4-1
Troubleshooting (Continued)
Symptom
Probable Cause
Repair Action
Содержание VHS-4
Страница 2: ...Copyright 2005 Vacuum Technologies VHS 4 VHS 6 and VHS 250 Diffusion Pumps ...
Страница 4: ...VHS 4 VHS 6 and VHS 250 Diffusion Pumps This page intentionally left blank ...
Страница 6: ...VHS 4 VHS 6 and VHS 250 Diffusion Pump This page intentionally left blank ...
Страница 8: ...VHS 4 VHS 6 and VHS 250 Diffusion Pumps This page intentionally left blank ...
Страница 10: ...VHS 4 VHS 6 and VHS 250 Diffusion Pumps This page intentionally left blank ...
Страница 12: ...VHS 4 VHS 6 and VHS 250 Diffusion Pumps This page intentionally left blank ...
Страница 34: ...VHS 4 VHS 6 and VHS 250 Diffusion Pumps This page intentionally left blank D R A F T 3 9 0 5 ...
Страница 50: ...VHS 4 VHS 6 and VHS 250 Diffusion Pumps This page intentionally left blank D R A F T 3 9 0 5 ...
Страница 56: ...VHS 4 VHS 6 and VHS 250 Diffusion Pumps This page intentionally left blank D R A F T 3 9 0 5 ...
Страница 59: ......