TECHNICAL INFORMATION
34
87-900-924-01(B)
N. Read/
Write
Data
Type
Description Admitted
Values
121 R/W N Maximum
rotational
frequency in
Hz (active
only in Stop
condition)
150 to 1050
(default = 1050)
122 R/W L Set
vent
valve on/off
(on =
closed)
On = 1
Off = 0
(default = 1)
123
124
Reserved to Varian service
125
R/W
L
Set the vent
valve
operation
Automatic = 0
(see note 1.)
On command = 1
(see note 2.)
126 R/W N Vent
valve
opening
delay
(expressed in
0.2 sec)
0 to 65535
(corresponding to
0 to 13107 sec)
130
Reserved to Varian service
200 R
(see
note 3.)
N Pump
current in
mA dc
0 to 500
201 R
(see
note 3.)
N Pump
volt-
age in Vac
0 to 400
202 R
(see
note 3.)
N Pump
power
in W (pump
current x
pump volt-
age x duty
cycle)
0 to 200
203 R N Driving
fre-
quency in
Hz
150 to 1050
204 R N Pump
tem-
perature in
°C
0 to 70
205
R
N
Pump status
Stop = 0
Waiting intlk = 1
Starting = 2
Regulation = 3
Braking = 4
Normal = 5
Fail = 6
206
R
N
Error code
Bit description:
no connection
(LSB), pump
overtempera-
ture, controller
overtempera-
ture, power fail,
auxiliary fail, over
voltage, short
circuit, too high
load (MSB)
210 R
(see
note 3.)
N Actual
rota-
tion speed
in Hz
150 to 1050
N. Read/
Write
Data
Type
Description Admitted
Values
300 R N Cycle
time
in minutes
(zeroed by
the reset
command)
0 to 4x10
9
301 R N Cycle
num-
ber (zeroed
by the reset
command)
0 to 4x10
9
302 R N Pump
life
in
hours (ze-
roed by the
reset com-
mand)
0 to 4x10
9
320
to
399
Reserved to Varian service
400 R A CRC
EPROM
(QE)
QE4XXXX
(where “XXXX”
are variable)
402 R A CRC
Pa-
ram. (PA)
PA4XXXX
(where “XXXX”
are variable)
500
Reserved to Varian service
503 R/W N Address
This
window
must be used
only when a sin-
gle controller is
connected to the
network
NOTES
1. Automatic means that when the controller
stops, the vent valve is opened with a delay de-
fined by window n. 126; when the controller
starts, the vent valve is immediately closed.
2. On command means that the vent valve is
opened or closed by means of window n. 122.
3. These commands remove the pump motor
supply for a short time. If they are used con-
tinuously, they can cause the pump stopping.
ACCESSORIES AND SPARE PARTS
DESCRIPTION PART
NUMBER
Fan 969-9323
Vent valve
969-9834
Purge valve with KF16 flange
969-9239
Purge valve with 7/16” flange
969-9240
RS485 cable
969-9881