background image

The quality of the power supplied to your Saturn GC/MS is very important. The power must be 
100-130 Vac, 60 Hz 

±

3 Hz (200-260 Vac, 50 Hz 

±

3 Hz outside North America), and it must be 

stable, i.e., it must be free of fluctuations due to slow changes in the average voltage or to 
changes resulting from surges, sags, or transients. 

  Slow average changes are gradual, long-term changes in the average root mean 

square (RMS) voltage level, with typical durations greater than 2 seconds. 

  Sags and surges are sudden changes in average RMS voltage level, with typical 

durations between 50 

µ

sec and 2 seconds. 

  Transients (or impulses) are brief voltage excursions of up to several thousand volts 

with durations of less than 50 

µ

sec. 

Constant high line voltage or surges in voltage may produce overheating and component 
failures. Constant low line voltage or sags in voltage may cause the system to function 
erratically, or even to cease functioning. Transients, even of a few microseconds duration, may 
cause electronic devices to fail catastrophically or degrade sufficiently to significantly shorten 
device lifetime. It is important to establish the quality of the line power in your laboratory prior to 
installation of the Saturn GC/MS. 

Occasionally, you may encounter line power sources of unacceptable quality; such power 
sources may adversely affect the operation of the Saturn GC/MS. You may want to contact a 
specialist in power conditioning services. 

Operating Environment 

You are responsible for providing an acceptable operating environment. Attention paid to the 
operating environment will ensure the continued peak performance of your Saturn GC/MS.  

Temperature/Humidity 

The optimum operating temperature is between 18

°

 and 21 

°

C (65

°

 and 70 

°

F). 

NOTE: As laboratory temperature increases, system reliability decreases. All electronic 
components generate heat while operating. This heat must be dissipated to the 
surrounding air if the components are to operate reliably. 

The turbomolecular pump temperature cutoff control protects the bearing and prolongs its 
lifetime. If the laboratory temperature is significantly above 27 

°

C (80 

°

F), the pump cutoff 

temperature could be reached; this would cause the pump to shut down. 

The diffusion pump’s temperature cutoff control protects the fluid from overheating. If the 
laboratory temperature is above 35 

°

C (95 

°

F), the pump will shut down. 

There must be a good flow of air around the system. Your air conditioning system must be 
capable of maintaining a constant temperature (within operational limits) in the immediate 
vicinity of the system. The average steady-state heat load of the Saturn GC/MS is 6000 BTUs, 
with a possible short-term heat dissipation of 15000 BTUs during startup. 

Hot air vented from the GC column oven may contribute to room heating and to the resulting air 
conditioning load. Ducting the GC column oven air out of the lab should reduce this heating 
effect.  

The relative humidity (RH) of the operating environment must be between 40 and 80%, with no 
condensation. Operating a Saturn GC/MS at very low humidity will result in the accumulation 
and discharge of static electricity shortening the life of electronic components. Operating the 
system at high humidity will produce condensation and result in short circuits.  

Varian recommends that your laboratory be equipped with a temperature/humidity monitor. This 
will ensure that your laboratory is always in conformance with temperature and humidity 
specifications. 

03-914629-00:7 

5

 

of 9 

Saturn 2000 GC/MS Pre-installation Instructions

 

Содержание Saturn 2000 GC/MS

Страница 1: ...tside of the power and operating environment limits described below could cause failures of many types The repair of such failures is specifically excluded from the Warranty and Service contract condi...

Страница 2: ...24 11 Tekmar 3000 Purge and Trap 19 48 9 23 18 46 37 17 Archon Purge and Trap AutoSampler 17 43 22 55 21 53 80 36 Tekmar ALS 2016 27 69 15 38 15 38 35 16 Genesis Headspace AutoSampler 22 56 28 71 18...

Страница 3: ...e fourplex power sources is for the mass spectrometer computer monitor and printer the other fourplex power source is for the gas chromatograph and optional 8200 AutoSampler If you have additional sam...

Страница 4: ...NEMA 5 15P plugs The power cable from the 8200 AutoSampler is about 2m 6 ft long and is fitted with a NEMA 5 15P plug rated at 120V With a 120V power source the maximum amperage requirements for each...

Страница 5: ...optimum operating temperature is between 18 and 21 C 65 and 70 F NOTE As laboratory temperature increases system reliability decreases All electronic components generate heat while operating This heat...

Страница 6: ...rmance of the Saturn GC MS it may also damage such components as the capillary column filaments and multiplier Varian recommends that you verify that your gas suppliers use controlled tanks this will...

Страница 7: ...he back panel of the CP 3800 GC or 3900 GC In addition the company network must allow TCP IP communications to be routed across it from the GC to the Saturn GC MS Workstation For basic Ethernet cablin...

Страница 8: ...5e driver Lexmark 3000 Color Jetprinter Inspection When your Saturn GC MS arrives carefully inspect the exterior of the shipping cartons for evidence of any damage that might have occurred during ship...

Страница 9: ...ans to analyze samples with the system until after the installation has been completed and you have accepted the conditions of delivery Parts and Supplies The Saturn 2000 GC MS Hardware Manual provide...

Отзывы: