ProStar 220/230/240 SDM Control Manual
-
17 -
03-914947-21:4
Use the following buttons to edit the table:
to add a line to the table.
to delete a line in the table.
to clear the entire table.
End of analysis behavior
:
In the scrolling list, select either to stop the pump or leave the
pump on at the end of the analysis (Quick Start or sequence).
If Hold is selected, the pump will continue to work at the last
time-table flow and composition.
If Equilibrate is selected, the pump will perform an equilibration
step to return to the initial conditions and will continue to work at
the initial conditions.
If Stop Pump (end of analysis) is selected, the pump will stop at
the end of each run.
If Stop Pump (end of sequence only), the pump will stop only
when the sequence is finished.
Enter in the corresponding fields the minimum and maximum
pressure limits. The pressure range is 0 to 410 Bar, (0 to 5946
PSI, and 0 to 40996 kPa).