Installation
1-15
c.
If the pump is hot, wait until the pump has cooled then vent it to atmosphere. Remove the
plug from the filling port of the pump and install the specified quantity of fluid up to the
FULL COLD level indicator on the sight glass (see Figure 1-8). The fluid can also be
poured in from the pump inlet or foreline.
d.
Replace the fill plug with a new Viton gasket. Tighten the fill plug using moderate torque
(75 inch-pounds maximum). Lubricate the O-ring with pump fluid before installation.
e.
Once the pump is running, the oil level should be in the FULL HOT operating range as
indicated on the sight glass assembly.
f.
It is not recommended that Santovac 5 Diffusion Pump Fluid be used with the NHS-35.
Figure 1-8. Sight Glass Assembly
Содержание F1730308
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 16: ...x ...
Страница 18: ......
Страница 20: ......
Страница 22: ...Installation 1 2 Figure 1 1 Speed and Throughput Curves NHS 35 Diffusion Pump ...
Страница 24: ...Installation 1 4 Figure 1 2 Physical Specifications NHS 35 Diffusion Pump with ASA Flanges optional ...
Страница 26: ...Installation 1 6 Figure 1 3 Section Drawing NHS 35 Diffusion Pump ...
Страница 30: ...Installation 1 10 Figure 1 5 240 400 440V Wiring Diagram NHS 35 Diffusion Pump ...
Страница 31: ...Installation 1 11 Figure 1 6 380 415V Wiring Diagram NHS 35 Diffusion Pump ...
Страница 32: ...Installation 1 12 Figure 1 7 480V Wiring Diagram NHS 35 Diffusion Pump ...
Страница 36: ...Installation 1 16 ...
Страница 48: ......