TECHNICAL INFORMATION
1. Disconnect the flange (28) from the inlet pipe line.
2. Remove the 6 securing screws (17) and remove the two half-plates (25) pressing the suction flange (28)
into its seat.
3. Remove the flange (28) and take out the piston (30) with the shutter (29) of the anti-suckback device, dis-
engage it by exerting force on arm (101). Separate the shutter (29) from the piston rod (30) simply by
pulling.
CAUTION
Be careful not to scratch the sealing surface of the shutter (29).
4. Reassemble by proceeding in the inverse order.