Reference 

 
 

External connections 

 
 

 

Digital IN 

 

 
 

 DIGITAL IN

GND

1

2

3

4

5

6

7

8

9

10

Pin

DIGITAL IN

External Ready-IN

Start Micro-GC externally

+12 VDC out ( electronic fused max 500 mA)

+5 VDC out (electronic fused max 500 mA)

Digital IN1 (not definied)

Digital IN2 (not definied)

External system reset

GND

External Opto input

(high is active)

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

The CP-4900 Micro-GC reacts, when “Digital IN” signals will be connected to Ground 
(GND).

Page 6   

User Manual CP-4900 Micro-GC Extension Box 

 

Varian, BV.  

Содержание CP-4900 Micro-GC

Страница 1: ...94598 California USA Tel 1 9259392400 Fax 1 9259452360 or 1 9259452344 Europe P O Box 8033 4330 EA Middelburg The Netherlands Tel 31 118671000 Fax 31 118623193 Australia East Asia 679 Springvale Road...

Страница 2: ...or accessory Repair or replacement under warranty does not extend the original warranty period Repair or replacement under warranty claims shall be made in Varian s sole discretion either by sending a...

Страница 3: ...h case the user will be required to correct the interference at his own expense To prevent any injury to the user or any damage to the instrument it is essential that you read the information in this...

Страница 4: ...ceeding 1000V must be so marked Indicates parts that may cause burns when touched Indicates that the user should refer to the manual before operating the equipment For protection against electrical sh...

Страница 5: ...l power sources before removing protective panels to avoid exposure to potentially dangerous voltages 2 When it is necessary to use a non original power cord plug make sure the replacement cord adhere...

Страница 6: ...ttempting any type of maintenance 14 Use proper tools when working on the instrument to prevent danger for you and or damage to the instrument 15 The customer should not attempt to replace any fuses i...

Страница 7: ...can be provided by qualified personnel on both a contractual or as needed basis Varian Analytical Instruments Sales Offices For Sales or Service assistance and to order Parts and Supplies contact your...

Страница 8: ......

Страница 9: ...ecautions 5 CAUTIONS 5 Spare Parts Availability 7 Service Availability 7 Table of contents 1 Introduction 1 Pre installation requirements 2 Environmental requirements 2 Space requirements 2 CP 4900 Mi...

Страница 10: ...Table of contents Shipping instructions 11 Cleaning instructions 11 Disposal instructions 11 Page 2 User Manual CP 4900 Micro GC Varian Inc...

Страница 11: ...on Box The CP 4900 Micro GC Extension Box is used to bring the CP 4900 Micro GC more additional in outputs For problems or questions about your CP 4900 Micro GC Extension Box please contact your neare...

Страница 12: ...ditioners heaters furnaces or fans Space requirements Allow sufficient bench space to permit installation of workstations integrators and other Micro GC equipment Allow 10 20 cm of space at the sides...

Страница 13: ...k the Micro GC CP 4900 Extension Box and accessories carefully and transfer to the work area using proper handling techniques Inspect the CP 4900 Micro GC Extension Box and accessories carefully for d...

Страница 14: ...Reference Cable connectors CD Rom Package contents slip Page 4 User Manual CP 4900 Micro GC Extension Box Varian BV...

Страница 15: ...GC use the included cables and connect according the picture below Always connect both cables Cable 25 pins Cable 15 pins Male Female Female Male CP 4900 Micro GC Extension Box C o n n e c t t o D I...

Страница 16: ...ly 12 VDC out electronic fused max 500 mA 5 VDC out electronic fused max 500 mA Digital IN1 not definied Digital IN2 not definied External system reset GND External Opto input high is active The CP 49...

Страница 17: ...rt GC relay 1 500 mS contact closure when injection take place This contact will follow instrument status Ready LED in front GC relay 2 Digital output 1 Digital output 2 All relay contacts maximum 1A...

Страница 18: ...6 7 8 Pin ANALOG IN Analog input 1 0 10 Volt Analog input 2 0 10 Volt Analog input 3 0 10 Volt Analog input 4 0 10 Volt Analog input 5 0 10 Volt Analog input 6 0 10 Volt AGND AGND Page 8 User Manual C...

Страница 19: ...Reference Relay 1 4 RELAY 1 4 1 2 3 4 5 6 7 8 Pin RELAY 1 4 Relay 1 Relay 2 Relay 3 Relay 4 All relay contacts maximum1A 24 VDC Varian BV User Manual CP 4900 Micro GC Extension Box Page 9...

Страница 20: ...Reference Relay 5 8 RELAY 5 8 1 2 3 4 5 6 7 8 Pin RELAY 5 8 Relay 5 Relay 6 Relay 7 Relay 8 All relay contacts maximum 1A 24 VDC Page 10 User Manual CP 4900 Micro GC Extension Box Varian BV...

Страница 21: ...fer to the remarks given below Clean only when CP 4900 Micro GC Extension Box is disconnected from the CP 4900 Micro GC or other equipment Use a soft no hard or abrasive brush to carefully brush away...

Отзывы: