CP-3800 GC Getting Started Manual
7
Configuration and Options
The 3800 has seven heated zones, one of which is always configured as the column
oven. The following options may be configured in the other six available heated
zones:
•
1079 Universal Capillary Injector
•
1041 On-Column Injector for large bore (0.53 mm ID) or packed columns
•
1061 Flash Vaporization Injector for large bore or packed columns
•
Single, dual, or multi-position Valve Oven
•
Flame Ionization Detector
•
Thermal Conductivity Detector
•
Electron Capture Detector
•
Thermionic Specific Detector
•
Pulsed Flame Photometric Detector
•
Methanizer
•
Micro Thermal Conductivity Detector
Any combination of three injectors may be installed on the 3800. If a valve oven is
installed, it normally replaces two injector positions.
Any combination of three detectors can be installed with the following exceptions:
Only two TCD or PFPD detectors can be installed. If two standard TCDs are
installed, then no other detectors can be accommodated. However, a dual TCD (two
cells in one housing) is available as a Custom Solution which allows installation of
another detector, such as an FID, behind it.
The methanizer typically occupies the same location as one of the detectors. Dual
8200 AutoSamplers restrict the 3800 configuration to two injectors and two detectors.
Installed options on the 3800 are generally identified by their location. The options
mounted on the top of the instrument, such as those listed above, are designated as
Front, Middle and Rear to coincide with their relative location or the location of their
respective electronic control modules.