Model 990 dCLD II Component Leak Detector
3-10
IDLE MODE
RUN MODE
IDLE mode allows the spectrometer tube and turbo pump to be
shut down without turning off main power. Volatile operating
parameters will not be lost in this mode. The factory default setting
is the IDLE function is disabled. The Idle function must be
ENABLED before it can be used.
To put the control unit in IDLE mode, de-assert the IDLE/RUN input
(J3/5) in the discrete I/O.
RUN mode powers the spectrometer tube and turbo pump. The
IDLE function must be ENABLED before RUN can be used.
To put the control unit in RUN mode, assert the IDLE/RUN input
(J3/5) in the discrete I/O
NOTE
The ENABLE-IDLE command must be issued for the IDLE and
RUN modes to be operational. The factory default setting is
DISABLE-IDLE.
CALIBRATE
CALIBRATE is used to perform an automated calibration based
upon the parameters set. Calibration is either to a known INTER-
NAL or EXTERNAL leak whose value has been entered into the
appropriate parameter, and is either performed in NORMAL or
FAST mode.
To perform an automated calibration, issue the CALIBRATE<CR>
command or, if the optional front control panel is installed, press
the CALIBRATE button.
CALIBRATED
LEAK VALUE
The 990 dCLD II is shipped from the factory with a calibrated leak
(Cal Leak) as part of the turbo spec tube assembly. Its value is noted
on the tag attached to it and on the Cal Leak Certification. During
setup, enter the value of the calibrated leak (Section 4.6.1
“Calibrated Leak Set-Up” on page 4-13). When a Calibration is per-
formed, the system compares the leak rate signal with the stored
value.
INTERNAL
To enter the value of the internal calibrated leak, issue the
INIT-STDLEAK<CR> command or, if the optional front control
panel is installed, enter the value via the Calibrated Leak Set-Up
screen (Section 4.6.4 “Range Stop and Manual Range Set Up and
Control” on page 4-15).
EXTERNAL
If you are using an external leak, you must enter its value so the
system can use it for comparison when a Calibration is performed.
To enter the value of the external calibrated leak, issue the
INIT-EXTLEAK<CR> command or, if the optional front control
panel is installed, enter the value via the Calibrated Leak Set-Up
screen (Section 4.6.1 “Calibrated Leak Set-Up” on page 4-13).
Содержание 990 dCLD II
Страница 4: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank D R A F T 4 2 0 0 7 ...
Страница 12: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 50: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 128: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 146: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 160: ...Model 990 dCLD II Component Leak Detector This page intentionally left blank ...
Страница 161: ......