Models 947, 948 and 960 Leak Detectors Operations and Service Manual
4-23
4.2.7 TC1 and TC2 Gauge Replacement and Calibration
4.2.7.1 Replacement of the TC1 Gauge in the Main Valve Block Assembly
Figure 4-10 shows the valve block removed from the system.
NOTE
It is not necessary to remove the valve block to replace the
TC1 thermocouple gauge.
To replace the TC1 gauge in the Main Valve Block Assembly (MVBA):
1. Press the VENT button on the front panel to vent the leak detector.
2. Disconnect the mains power supply and make sure all pumps are off.
3. Grasp the plug and remove the plug and the thermocouple gauge
➀
from the valve
4. Disconnect the plug from the thermocouple gauge.
5. Remove the O-ring
➂
and backing ring
➁
from the thermocouple gauge (Figure 4-10).
Figure 4-10
TC1 Gauge in the MVBA
➀
➁
➂
Содержание 948
Страница 8: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 36: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 102: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 144: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 152: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 162: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 175: ......