Models 947, 948 and 960 Leak Detectors Operations and Service Manual
4-9
4.2.2.2 Replacing the Internal Calibrated Leak
To replace the calibrated leak:
1. Wipe the O-ring with a lint-free cloth to ensure that no foreign matter is present.
CAUTION
Varian recommends replacing all used O-rings.
Do not use grease or other substance on O-rings that come in
contact with the spectrometer tube.
If applicable, apply a small amount of Apiezon
L grease and
wipe the O-rings “shiny” dry.
Do not use alcohol, methanol or other solvents on O-rings. To
do so causes deterioration and reduces their ability to hold a
vacuum.
2. Place the O-ring onto the centering ring.
3. Place the centering ring onto the calibrated leak.
4. Insert the calibrated leak into the cavity, orienting the leak so the value can be
read easily.
NOTE
To ensure reliable testing, after servicing any part of the
vacuum system, the 947/948/960 must be leak checked again.
Refer to “Leak Checking” on page 4-5 for guidelines.
5. Replace the top panel and connect mains power.
6. Turn the leak detector on and wait for the pumps to come up to speed.
7.
Models 948/960 only:
Once the system has pumped down, clean and fill the
Liquid Nitrogen Cold Trap using the procedure described in “Liquid Nitrogen Cold Trap
(Models 948/960 only)” on page 4-10.
Follow all warnings and cautions in the procedure.
8. Calibrate the 947/948/960 using the procedure described in “Re-calibrate the Internal
9. Leak check the assembly before using the 947/948/960 for testing.
Содержание 948
Страница 8: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 36: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 102: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 144: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 152: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 162: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 175: ......