Models 947, 948 and 960 Leak Detectors Operations and Service Manual
B-6
B.2 Replacing the Ion Source
If the ion source needs replacement (Figure B-5), replace it during spectrometer tube
cleaning. Adjustment is required after replacement is complete. Refer to
Sensitivity Checks” on page 3-11 for the adjustment procedures.
WARNING
Store the Ion Source in a cool, dry area in a tightly sealed
container. Wash hands thoroughly after handling the Ion Source
and especially before smoking or eating.
CAUTION
Use powder-free butyl or polycarbonate gloves to prevent skin
oils from getting on vacuum surfaces.
To replace the ion source:
1. Remove the spectrometer tube from the leak detector as described in “Removing and
Replacing both the Spectrometer Tube Assembly and Preamplifier” on page 4-21.
Figure B-5
Ion Source Replacement
A
A
➁
➂
➄
➅
➄
➃
➀
Содержание 948
Страница 8: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 36: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 102: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 144: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 152: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 162: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank...
Страница 175: ......