Models 947, 948 and 960 Leak Detectors Operations and Service Manual
3-11
Refill the cold trap as required to prevent it from thawing in the vacuum system. The
trap needs to be filled approximately every 14 hours. Monitor the level and fill more
often if necessary.
CAUTION
If the cold trap in the 948/960 is allowed to empty of
liquid and thaw in the vacuum system, spectrometer
tube contamination occurs resulting in unstable leak
rate detection. Attempts to perform an automated
calibration can also result in error conditions (codes c6,
c7, and c8). Filling the trap completely keeps the trap
operational up to 14 hours. Before refilling is required,
refer to Section 4, “Maintenance” for the cold trap
cleaning procedure.
5. Wait 30 minutes for the 947/948/960 to warm up to operating temperature.
6. The detector is now on and ready for normal operation.
NOTE
Perform the calibration and sensitivity check before
taking readings with the 947/948/960.
3.4 Calibration and Sensitivity Checks
At the beginning of each workday or shift, or as conditions require, verify 947/948/960
integrity by performing the Calibration and Sensitivity Check procedure. Compliance with
this procedure prepare your detector for normal daily operation. If the
NOT READY
LED
comes on during calibration, refer to the condition code that appears in the leak rate
exponent display shown in Figure 3-3.
Figure 3-3
Exponent Display with Condition Code Indicated
➀
Condition Code
Exponent area with condition code displayed.
➀
Содержание 947
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 16: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 80: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 120: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 160: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual 4 40 ...
Страница 176: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 206: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 214: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 228: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 241: ......