B-16
B.4.1.1 Removing the Wet System Rough Pump
To perform this procedure:
1. Disconnect the mains power supply and make sure that all pumps are off.
WARNING
Disconnect power from the 947/948/960 before
performing any maintenance procedure that requires
physically disconnecting any part of the system.
CAUTION
Models 948/960 only:
Remove, empty, and clean the
cold trap whenever power is removed. Refer to “Liquid
Nitrogen Cold Trap (Models 948/960 only)” on
page 4-10.
2. Remove the front panel.
3. Remove the right side panel.
4. Disconnect the pump power plug from the 947/948/960.
5. Drain the oil according to the pump manual instructions.
6. If you are replacing the pump as part of a Varian exchange program, return it in
accordance with the program, or discard it in accordance with all applicable
regulations.
NOTE
Varian Vacuum Technologies will not accept any part
for return without both a properly completed Return
Materials Authorization (RMA) and a Health and Safety
Data Sheet. Contact Customer Service at
1-800-8VARIAN to obtain an RMA and a Health and
Safety Data Sheet.
7. Remove the NW25 clamp from the left manifold (Figure B-11).
Figure B-11
Rough Pump Manifold and Exhaust Connections
Exhaust Connection
Manifold Connection
NW25 Clamp
NW16 Clamp
Содержание 947
Страница 12: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 16: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 80: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 120: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 160: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual 4 40 ...
Страница 176: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 206: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 214: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 228: ...Models 947 948 and 960 Leak Detectors Operations and Service Manual This page intentionally left blank ...
Страница 241: ......