XRpad2 4343
12
www.vareximaging.com
U
SER
M
ANUAL
10.3
X-ray Detector Specification
displays the specifications for the X-ray detector. In addition,
, and
below display the measured DQE (per the IEC 62220-1 Ed.1 standard at ~8 µGy), MTF,
and characteristic curves of the X-ray detector.
Table 7
X-ray Detector Specification
Sensor
Panel
Single substrate amorphous silicon active TFT/diode array
Scintillator
Direct deposition CsI:Tl
Gd
2
O
2
S:Tb (Gadox)
Pixel Matrix
4288 × 4288
Pixel Pitch
100
μ
m
Electronics
Amplifiers
Low noise ASICs with user selectable gains
ADC
16-bit
Image Transfer Time
Wired: 1000 ms; Wireless: 5500 ms
On-board Memory
1 GB DDR3, 8 GB SDHC card
Mechanical
Size
ISO 4090 for 43 cm × 43 cm (17” × 17”) cassette size
Active Area
426 mm × 426 mm
External Dimensions
460 mm (w) × 460 mm (l) × 15.5 mm (h)
Weight
4.1 kg (9.03 lbs)
Housing
Aluminum frame with carbon-fiber entrance window
Communication
Status Display
OLED display with Wi-Fi, LAN, battery, and sensor indicators
Wireless Data I/F
802.11n Wi-Fi standard @ 5 GHz
Channel: WAP:
36-48, 149-165
Stationary:
36-48, 149-165
Stationary DFS:
a
52-64, 100-116, 132-140
Wired Data I/F
GigE, trigger and power via optional docking connector
X-ray I/F
Integrated X-ray trigger control
Automatic Exposure Detection
Image Performance
Limiting Resolution
5.0 cy/mm
Typical MTF
70% (1 cy/mm), 40% (2 cy/mm), 15% (4 cy/mm) for RQA5
Typical DQE
75% (0 cy/mm), 60% (1 cy/mm), 40% (3 cy/mm) for RQA5
Typical Radiographic Lag
1%
Version
01
2017-11-03